Effective Radiation Dose from Cone-Beam Computed Tomography Guidance during Bronchoscopic Tumour Ablation.
Respiration
; 103(8): 498-502, 2024.
Article
em En
| MEDLINE
| ID: mdl-38885628
ABSTRACT
INTRODUCTION:
Endobronchial radiofrequency ablation (RFA) is a novel minimally invasive approach to management of peripheral non-small-cell lung cancer (NSCLC) in medically inoperable patients. Minimally invasive ablative techniques are generally delivered with cone-beam computed tomography (CBCT) guidance. CBCT requires a significant number of two dimensional imaging projections to be acquired which is then reconstructed as a three-dimensional cone-beam image. The objective of this study was to determine the radiation dosimetry consequent to use of CBCT guidance for bronchoscopic RFA.METHODS:
Post hoc analysis of data following bronchoscopic RFA of stage I biopsy-confirmed NSCLC performed with CBCT. Effective dose estimates for these patients were calculated using PCXMC2.0 software.RESULTS:
Ten patients underwent bronchoscopic RFA, with a median 3 (range 2-4) CBCT spins per procedure. Mean dose area product (DAP) per procedure was 7,778 µGy.m2 (±4,743) with an effective dose of 11.6 mSv (±7.4). The DAP per spin for these 10 patients varied from 83.8 to 8,625.6 µGy.m2 (effective dose range 0.15-13.81 mSv).CONCLUSION:
This is the first study to report radiation dosimetry consequent to CT guidance for bronchoscopic RFA procedures. Effective doses appear comparable to other CT fluoroscopic procedures.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Doses de Radiação
/
Broncoscopia
/
Carcinoma Pulmonar de Células não Pequenas
/
Tomografia Computadorizada de Feixe Cônico
/
Neoplasias Pulmonares
Limite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Ano de publicação:
2024
Tipo de documento:
Article