Your browser doesn't support javascript.
loading
High-Throughput Proteomics Enabled by a Fully Automated Dual-Trap and Dual-Column LC-MS.
Chen, Liang; Zhang, Ziwei; Matsumoto, Cory; Gao, Yu.
Afiliação
  • Chen L; College of Pharmacy, University of Illinois at Chicago, Chicago, Illinois 60612, United States.
  • Zhang Z; College of Pharmacy, University of Illinois at Chicago, Chicago, Illinois 60612, United States.
  • Matsumoto C; College of Pharmacy, University of Illinois at Chicago, Chicago, Illinois 60612, United States.
  • Gao Y; College of Pharmacy, University of Illinois at Chicago, Chicago, Illinois 60612, United States.
Anal Chem ; 96(24): 9761-9766, 2024 06 18.
Article em En | MEDLINE | ID: mdl-38887087
ABSTRACT
This Technical Note describes a dual-column liquid chromatography system coupled to mass spectrometry (LC-MS) for high-throughput bottom-up proteomic analysis. This system made full use of two 2-position 10-port valves and a binary pump with an integrated loading pump of a commercial LC instrument to provide successive operation of two parallel subsystems. Each subsystem consisted of a set of trap columns and an analytical column. A T-junction union was used to split the mobile phase from the loading pump into two parts. This allowed one set of columns to be washed and equilibrated, followed by the injection of the next sample, while the previous sample was eluting and being analyzed on the other set of columns, thereby greatly increasing the analysis throughput. This approach showed high reproducibility for the analysis of HeLa tryptic digests with average relative standard deviation (RSD) values of 1.75%, 6.90%, and 5.19% for the identification number of proteins, peptides, and peptide-spectrum matches (PSMs), respectively, across 10 consecutive runs. The capacity for peptide and protein identification, as well as proteome depth, of the dual-column LC system was comparable to a conventional single-column system. Due to its simple equipment requirements and set up process, this method should be highly accessible for other laboratories.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Proteômica Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Espectrometria de Massas / Proteômica Limite: Humans Idioma: En Ano de publicação: 2024 Tipo de documento: Article