Your browser doesn't support javascript.
loading
A preliminary study of super-resolution deep learning reconstruction with cardiac option for evaluation of endovascular-treated intracranial aneurysms.
Otgonbaatar, Chuluunbaatar; Kim, Hyunjung; Jeon, Pil-Hyun; Jeon, Sang-Hyun; Cha, Sung-Jin; Ryu, Jae-Kyun; Jung, Won Beom; Shim, Hackjoon; Ko, Sung Min; Kim, Jin Woo.
Afiliação
  • Otgonbaatar C; Department of Radiology, College of Medicine, Seoul National University, Seoul, 03080, Republic of Korea.
  • Kim H; Medical Imaging AI Research Center, Canon Medical Systems Korea, Seoul, 06173, Republic of Korea.
  • Jeon PH; Department of Radiology, Wonju Severance Christian Hospital, Wonju College of Medicine, Yonsei University of Korea, Wonju 26426, Republic of Korea.
  • Jeon SH; Department of Radiology, Wonju Severance Christian Hospital, Wonju College of Medicine, Yonsei University of Korea, Wonju 26426, Republic of Korea.
  • Cha SJ; Department of Radiology, Wonju Severance Christian Hospital, Wonju College of Medicine, Yonsei University of Korea, Wonju 26426, Republic of Korea.
  • Ryu JK; Department of Radiology, Wonju Severance Christian Hospital, Wonju College of Medicine, Yonsei University of Korea, Wonju 26426, Republic of Korea.
  • Jung WB; Medical Imaging AI Research Center, Canon Medical Systems Korea, Seoul, 06173, Republic of Korea.
  • Shim H; Korea Brain Research Institute (KBRI), Daegu, 41062, Republic of Korea.
  • Ko SM; Medical Imaging AI Research Center, Canon Medical Systems Korea, Seoul, 06173, Republic of Korea.
  • Kim JW; ConnectAI Research Center, Yonsei University College of Medicine, Seoul, 03772, Republic of Korea.
Br J Radiol ; 97(1160): 1492-1500, 2024 Aug 01.
Article em En | MEDLINE | ID: mdl-38917414
ABSTRACT

OBJECTIVES:

To investigate the usefulness of super-resolution deep learning reconstruction (SR-DLR) with cardiac option in the assessment of image quality in patients with stent-assisted coil embolization, coil embolization, and flow-diverting stent placement compared with other image reconstructions.

METHODS:

This single-centre retrospective study included 50 patients (mean age, 59 years; range, 44-81 years; 13 men) who were treated with stent-assisted coil embolization, coil embolization, and flow-diverting stent placement between January and July 2023. The images were reconstructed using filtered back projection (FBP), hybrid iterative reconstruction (IR), and SR-DLR. The objective image analysis included image noise in the Hounsfield unit (HU), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and full width at half maximum (FWHM). Subjectively, two radiologists evaluated the overall image quality for the visualization of the flow-diverting stent, coil, and stent.

RESULTS:

The image noise in HU in SR-DLR was 6.99 ± 1.49, which was significantly lower than that in images reconstructed with FBP (12.32 ± 3.01) and hybrid IR (8.63 ± 2.12) (P < .001). Both the mean SNR and CNR were significantly higher in SR-DLR than in FBP and hybrid IR (P < .001 and P < .001). The FWHMs for the stent (P < .004), flow-diverting stent (P < .001), and coil (P < .001) were significantly lower in SR-DLR than in FBP and hybrid IR. The subjective visual scores were significantly higher in SR-DLR than in other image reconstructions (P < .001).

CONCLUSIONS:

SR-DLR with cardiac option is useful for follow-up imaging in stent-assisted coil embolization and flow-diverting stent placement in terms of lower image noise, higher SNR and CNR, superior subjective image analysis, and less blooming artifact than other image reconstructions. ADVANCES IN KNOWLEDGE SR-DLR with cardiac option allows better visualization of the peripheral and smaller cerebral arteries. SR-DLR with cardiac option can be beneficial for CT imaging of stent-assisted coil embolization and flow-diverting stent.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Stents / Aneurisma Intracraniano / Procedimentos Endovasculares / Aprendizado Profundo Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Stents / Aneurisma Intracraniano / Procedimentos Endovasculares / Aprendizado Profundo Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2024 Tipo de documento: Article