Your browser doesn't support javascript.
loading
Values of peripapillary retinal nerve fibre layer thickness are different in children and adults.
Muñoz-Gallego, Alicia; Torres-Peña, José L; Rodríguez-Salgado, Martín; Ortueta-Olartecoechea, Ana; López-López, Cristina; De La Cruz, Javier; Tejada-Palacios, Pilar.
Afiliação
  • Muñoz-Gallego A; Ophthalmology Department, Madrid University Complutense, Madrid, Spain.
  • Torres-Peña JL; Ophthalmology Department, Madrid University Complutense, Madrid, Spain.
  • Rodríguez-Salgado M; Madrid Regional Health Service (SERMAS), Primary Health Care, Madrid, Spain.
  • Ortueta-Olartecoechea A; Ophthalmology Department, Madrid University Complutense, Madrid, Spain.
  • López-López C; Ophthalmology Department, Madrid University Complutense, Madrid, Spain.
  • De La Cruz J; Research Institute (imas12), Madrid University Hospital, Madrid, Spain.
  • Tejada-Palacios P; Ophthalmology Department, Madrid University Complutense, Madrid, Spain.
Clin Exp Optom ; : 1-9, 2024 Jul 18.
Article em En | MEDLINE | ID: mdl-39025791
ABSTRACT
CLINICAL RELEVANCE The detection of abnormal values of peripapillary nerve fibre layer (pRNFL) thickness measured with optical coherence tomography (OCT) is important for detecting optic nerve disease in children.

BACKGROUND:

To evaluate the level of agreement between the adult reference database supplied with an OCT device and the present paediatric study database for the measurement of pRNFL thickness in children. This study also aimed to provide reference values for pRNFL thickness according to the spherical equivalent in the paediatric population.

METHODS:

This was a cross-sectional study. One hundred and twenty-six healthy children were included, who had undergone a full ophthalmological examination including cycloplegic refraction and examination of pRNFL thickness using the Topcon 3D OCT 2000 device (Topcon Corporation, Tokyo, Japan). Values equal to or below the fifth percentile (≤p5) and above the 95th percentile (>p95) were considered abnormal. Observed agreement and specific agreement were investigated between OCT measurements classified with paediatric and adult reference values for normality.

RESULTS:

Values ≤ p5 in the adult database were recorded for 2 of the 30 values (6.6%) of the pRNFL values by quadrants ≤p5 in the paediatric database and 17 of the 88 (19.3%) values by sectors ≤p5. For values >p95 in the adult database, 88% by quadrants and 72% by sectors would have been classified as being within the normal range using the paediatric database.

CONCLUSION:

The use of adult reference values currently available in OCT devices can lead to classification errors concerning the normal range of pRNFL thickness in a large proportion of paediatric patients. The use of normative paediatric databases, such as the one discussed in this study, should be taken into consideration.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article