Fully automatic estimation of global left ventricular systolic function using deep learning in transoesophageal echocardiography.
Berg, Erik Andreas Rye; Taskén, Anders Austlid; Nordal, Trym; Grenne, Bjørnar; Espeland, Torvald; Kirkeby-Garstad, Idar; Dalen, Håvard; Holte, Espen; Stølen, Stian; Aakhus, Svend; Kiss, Gabriel.
Afiliação
  • Berg EAR; Centre for Innovative Ultrasound Solutions, Department of Circulation and Medical Imaging, Faculty of Medicine and Health Science, Norwegian University of Science and Technology, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Taskén AA; Department of Circulation and Medical Imaging, Faculty of Medicine and Health Science, Norwegian University of Science and Technology, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Nordal T; Clinic of Cardiology, St Olavs Hospital, Trondheim University Hospital, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Grenne B; Department of Computer Science, Faculty of Information Technology and Electrical Engineering, Norwegian University of Science and Technology, Trondheim 7034, Norway.
  • Espeland T; Department of Engineering Cybernetics, Faculty of Information Technology and Electrical Engineering, Norwegian University of Science and Technology, Trondheim 7034, Norway.
  • Kirkeby-Garstad I; Department of Circulation and Medical Imaging, Faculty of Medicine and Health Science, Norwegian University of Science and Technology, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Dalen H; Clinic of Cardiology, St Olavs Hospital, Trondheim University Hospital, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Holte E; Department of Circulation and Medical Imaging, Faculty of Medicine and Health Science, Norwegian University of Science and Technology, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Stølen S; Clinic of Cardiology, St Olavs Hospital, Trondheim University Hospital, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Aakhus S; Department of Circulation and Medical Imaging, Faculty of Medicine and Health Science, Norwegian University of Science and Technology, Prinsesse Kristinas gate 3, Trondheim 7030, Norway.
  • Kiss G; Department of Anaesthesiology and Intensive Care Medicine, St Olavs Hospital, Trondheim University Hospital, Trondheim 7030, Norway.
Eur Heart J Imaging Methods Pract ; 1(1): qyad007, 2023 May.
Article em En | MEDLINE | ID: mdl-39044786
ABSTRACT

Aims:

To improve monitoring of cardiac function during major surgery and intensive care, we have developed a method for fully automatic estimation of mitral annular plane systolic excursion (auto-MAPSE) using deep learning in transoesophageal echocardiography (TOE). The aim of this study was a clinical validation of auto-MAPSE in patients with heart disease. Methods and

results:

TOE recordings were collected from 185 consecutive patients without selection on image quality. Deep-learning-based auto-MAPSE was trained and optimized from 105 patient recordings. We assessed auto-MAPSE feasibility, and agreement and inter-rater reliability with manual reference in 80 patients with and without electrocardiogram (ECG) tracings. Mean processing time for auto-MAPSE was 0.3 s per cardiac cycle/view. Overall feasibility was >90% for manual MAPSE and ECG-enabled auto-MAPSE and 82% for ECG-disabled auto-MAPSE. Feasibility in at least two walls was ≥95% for all methods. Compared with manual reference, bias [95% limits of agreement (LoA)] was -0.5 [-4.0, 3.1] mm for ECG-enabled auto-MAPSE and -0.2 [-4.2, 3.6] mm for ECG-disabled auto-MAPSE. Intra-class correlation coefficient (ICC) for consistency was 0.90 and 0.88, respectively. Manual inter-observer bias [95% LoA] was -0.9 [-4.7, 3.0] mm, and ICC was 0.86.

Conclusion:

Auto-MAPSE was fast and highly feasible. Inter-rater reliability between auto-MAPSE and manual reference was good. Agreement between auto-MAPSE and manual reference did not differ from manual inter-observer agreement. As the principal advantages of deep-learning-based assessment are speed and reproducibility, auto-MAPSE has the potential to improve real-time monitoring of left ventricular function. This should be investigated in relevant clinical settings.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2023 Tipo de documento: Article