A comparison between IBEX bone health applied to digital radiographs and dual-energy X-ray absorptiometry at the distal-third and ultra-distal regions of the radius.
Lopez, Ben; Meertens, Robert; Gundry, Mike; Scott, Paul; Crone, Mr Ben; McWilliam, Richard.
Afiliação
  • Lopez B; Research and Development, Ibex Innovations Ltd, Explorer 2, Netpark, Sedgefield, County Durham, TS21 3FF, UK. b.lopez@ibexinnovations.co.uk.
  • Meertens R; St Luke's Campus, University of Exeter, Exeter, EX1 2LU, UK.
  • Gundry M; St Luke's Campus, University of Exeter, Exeter, EX1 2LU, UK.
  • Scott P; Research and Development, Ibex Innovations Ltd, Explorer 2, Netpark, Sedgefield, County Durham, TS21 3FF, UK.
  • Crone MB; Research and Development, Ibex Innovations Ltd, Explorer 2, Netpark, Sedgefield, County Durham, TS21 3FF, UK.
  • McWilliam R; Research and Development, Ibex Innovations Ltd, Explorer 2, Netpark, Sedgefield, County Durham, TS21 3FF, UK.
BMC Musculoskelet Disord ; 25(1): 575, 2024 Jul 24.
Article em En | MEDLINE | ID: mdl-39049030
ABSTRACT

BACKGROUND:

In an ageing population, low impact fragility fractures are becoming increasingly common. However, fracture risk can be reduced where low bone density can be identified at an early stage. In this study we aim to demonstrate that IBEX Bone Health (IBEX BH) can provide a clinically useful prediction from wrist radiographs of aBMD and T-score at the ultra-distal (UD) and distal-third (DT) regions of the radius.

METHODS:

A 261-participant single-centre, non-randomised, prospective, study was carried out to compare a) IBEX BH, a quantitative digital radiography software device, to b) Dual-energy X-ray Absorptiometry (DXA). A total of 257 participants with wrist digital radiograph (DR), forearm DXA pairs were included in the analysis after exclusions.

RESULTS:

The adjusted R2 value for IBEX BH outputs to the radial areal bone mineral density (aBMD) produced by a GE Lunar DXA system for the UD region is 0.87 (99% Confidence Interval (CI) [0.84, 0.89]). The adjusted R2 value for IBEX BH outputs to aBMD for the DT region is 0.88 (99% CI [0.85, 0.90]). The Area Under the Receiver Operating Characteristic curve (AUC) for the forearm T-score ≤ - 2.5 risk prediction model at the UD region is 0.95 (99% CI [0.93, 0.98]). The AUC for the forearm T-score ≤ - 2.5 risk prediction model at the DT region is 0.98 (99% CI [0.97, 0.99]).

CONCLUSION:

From a DR of the wrist, IBEX BH provides a clinically useful i) estimate of aBMD at the two regions of interest on the radius and ii) risk prediction model of forearm T-score ≤ - 2.5 at the UD and DT regions.