Your browser doesn't support javascript.
loading
Raman spectroscopy as an alternative rapid microbial bioburden test method for continuous, automated detection of contamination in biopharmaceutical drug substance manufacturing.
Masucci, Erin M; Hauschild, James E; Gisler, Helena M; Lester, Erin M; Balss, Karin M.
Afiliação
  • Masucci EM; Janssen Pharmaceuticals Inc.  Welsh and McKean Roads, Spring House, PA 19477.
  • Hauschild JE; Johnson & Johnson Services, Inc.  1000 Route 202 South, Building 930 East, Raritan, NJ 08869.
  • Gisler HM; Janssen Pharmaceuticals Inc.  Welsh and McKean Roads, Spring House, PA 19477.
  • Lester EM; Janssen Pharmaceuticals Inc.  Welsh and McKean Roads, Spring House, PA 19477.
  • Balss KM; Janssen Pharmaceuticals Inc.  Welsh and McKean Roads, Spring House, PA 19477.
J Appl Microbiol ; 2024 Jul 25.
Article em En | MEDLINE | ID: mdl-39054049
ABSTRACT

AIMS:

To investigate an in-line Raman method capable of detecting accidental microbial contamination in pharmaceutical vessels, such as bioreactors producing monoclonal antibodies via cell culture. METHODS AND

RESULTS:

The Raman method consists of a multivariate model built from Raman spectra collected in-line during reduced-scale bioreactor batches producing a monoclonal antibody, as well as a reduced-scale process with intentional spiking of representative compendial method microorganisms (n = 4). The orthogonal partial least squares regression discriminant analysis model (OPLS-DA) Area Under the Curve (AUC), specificity and sensitivity were 0.96, 0.99 and 0.95, respectively. Furthermore, the model successfully detected contamination in an accidentally contaminated manufacturing-scale batch. In all cases, the time to detection (TTD) for Raman was superior compared to offline, traditional microbiological culturing.

CONCLUSIONS:

The Raman OPLS-DA method met acceptance criteria for equivalent decision making to be considered a viable alternative to the compendial method for in-process bioburden testing. The in-line method is automated, non-destructive, and provides a continuous assessment of bioburden compared to an offline compendial method, which is manual, results in loss of product, and in practice is only collected once daily and requires 3-5 days for enumeration.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article