Your browser doesn't support javascript.
loading
The effects of aperture position and length in side-vented needles on root canal irrigation: A computational fluid dynamics study.
Dai, Yuhua; Wang, Shizhao; Gao, Jiayang; Liu, Qi; Xiong, Zhe; Gao, Yuan.
Afiliação
  • Dai Y; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, Department of Cariology and Endodontics, West China Hospital of Stomatology, Sichuan University, Chengdu, China. Electronic address: good.fortune.denise@gmail.com.
  • Wang S; The State Key Laboratory of Nonlinear Mechanics, Institute of Mechanics, Chinese Academy of Sciences, Beijing, China; School of Engineering Sciences, University of Chinese Academy of Sciences, Beijing, China. Electronic address: wangsz@lnm.imech.ac.cn.
  • Gao J; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, Department of Cariology and Endodontics, West China Hospital of Stomatology, Sichuan University, Chengdu, China. Electronic address: gaojy22@163.com.
  • Liu Q; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, Department of Cariology and Endodontics, West China Hospital of Stomatology, Sichuan University, Chengdu, China. Electronic address: liuqi_0429@163.com.
  • Xiong Z; Clinical Medical College of Tianjin Medical University, Tianjin, China. Electronic address: 568351717@139.com.
  • Gao Y; State Key Laboratory of Oral Diseases, National Clinical Research Center for Oral Diseases, Department of Cariology and Endodontics, West China Hospital of Stomatology, Sichuan University, Chengdu, China. Electronic address: gaoyuan@scu.edu.cn.
Comput Biol Med ; 179: 108940, 2024 Sep.
Article em En | MEDLINE | ID: mdl-39059213
ABSTRACT

INTRODUCTION:

Root canal irrigation is crucial for infection control during root canal treatment. Side-vented needles for positive pressure irrigation are commonly used in clinical practice. However, variations in needle design among manufacturers can impact the fluid dynamics of irrigation. This study aims to use computational fluid dynamics to explore the flow characteristics of different needle aperture lengths and positions, and their effects on the effectiveness and safety of irrigation, using a validated passive scalar transport numerical model.

METHODS:

The validation of the CFD irrigant model was achieved by comparing it with an in vitro irrigation experiment model. The CFD model used scalar concentration, while the in vitro experiment model used red dye tracing. Using a standard 30G side-vented needle as a reference, virtual needle models featuring four aperture lengths and three positions were created. These virtual irrigation needles were then placed in two root canal geometries for CFD simulation to evaluate fluid exchange capabilities and related fluid dynamic parameters.

RESULTS:

The results of the CFD simulation, using a scalar transport model, closely matched the in vitro tracer tests for irrigation experiments across seven root canal geometries. The CFD analysis indicated that positioning the aperture lower increased the irrigant exchange distance. Notably, decreasing the aperture length to 0.25x, and positioning it at the lower end of the needle significantly increased exchange distance and shear stress, while reducing apical pressure.

CONCLUSIONS:

These results indicate that the position and length of the aperture affect the exchange distance of irrigant flow, wall shear stress, and apical pressure. The CFD validation model for scalar transport, based on a steady state, can function as a valuable tool for optimizing the side-vented needle in research. Further research on the design of side-vented needles will enhance the understanding of flow characteristics beneficial for irrigation efficiency in clinical practice.