Free-breathing three-dimensional simultaneous myocardial T1 and T2 mapping based on multi-parametric SAturation-recovery and Variable-flip-Angle.
Si, Dongyue; Guo, Rui; Cheng, Lan; Kong, Xiangchuang; Herzka, Daniel A; Ding, Haiyan.
Afiliação
  • Si D; Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China.
  • Guo R; School of Medical Technology, Beijing Institute of Technology, Beijing, China.
  • Cheng L; Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China; Hubei Province Key Laboratory of Molecular Imaging, Wuhan, China.
  • Kong X; Department of Radiology, Union Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, China; Hubei Province Key Laboratory of Molecular Imaging, Wuhan, China. Electronic address: hongke80@hust.edu.cn.
  • Herzka DA; Department of Radiology, Case Western Reserve University, Cleveland, Ohio, USA.
  • Ding H; Center for Biomedical Imaging Research, Department of Biomedical Engineering, School of Medicine, Tsinghua University, Beijing, China. Electronic address: dinghy@mail.tsinghua.edu.cn.
J Cardiovasc Magn Reson ; 26(2): 101065, 2024 Jul 24.
Article em En | MEDLINE | ID: mdl-39059610
ABSTRACT

BACKGROUND:

Quantitative myocardial tissue characterization with T1 and T2 parametric mapping can provide an accurate and complete assessment of tissue abnormalities across a broad range of cardiomyopathies. However, current clinical T1 and T2 mapping tools rely predominantly on two-dimensional (2D) breath-hold sequences. Clinical adoption of three-dimensional (3D) techniques is limited by long scan duration. The aim of this study is to develop and validate a time-efficient 3D free-breathing simultaneous T1 and T2 mapping sequence using multi-parametric SAturation-recovery and Variable-flip-Angle (mSAVA).

METHODS:

mSAVA acquires four volumes for simultaneous whole-heart T1 and T2 mapping. We validated mSAVA using simulations, phantoms, and in-vivo experiments at 3T in 11 healthy subjects and 11 patients with diverse cardiomyopathies. T1 and T2 values by mSAVA were compared with modified Look-Locker inversion recovery (MOLLI) and gradient and spin echo (GraSE), respectively. The clinical performance of mSAVA was evaluated against late gadolinium enhancement (LGE) imaging in patients.

RESULTS:

Phantom T1 and T2 by mSAVA showed a strong correlation to reference sequences (R2 = 0.98 and 0.99). In-vivo imaging with an imaging resolution of 1.5 × 1.5 × 8 mm3 could be achieved. Myocardial T1 and T2 of healthy subjects by mSAVA were 1310 ± 46 and 44.6 ± 2.0 ms, respectively, with T1 standard deviation higher than MOLLI (105 ± 12 vs 60 ± 16 ms) and T2 standard deviation lower than GraSE (4.5 ± 0.8 vs 5.5 ± 1.0 ms). mSAVA T1 and T2 maps presented consistent findings in patients undergoing LGE. Myocardial T1 and T2 of all patients by mSAVA were 1421 ± 79 and 47.2 ± 3.3 ms, respectively.

CONCLUSION:

mSAVA is a fast 3D technique promising for clinical whole-heart T1 and T2 mapping.
Palavras-chave