Day-to-Day Variability in Measurements of Respiration Using Bioimpedance from a Non-Standard Location.
Goyal, Krittika; Shah, Dishant; Day, Steven W.
Afiliação
  • Goyal K; Department of Manufacturing and Mechanical Engineering Technology, Rochester Institute of Technology, Rochester, NY 14623, USA.
  • Shah D; Department of Manufacturing and Mechanical Engineering Technology, Rochester Institute of Technology, Rochester, NY 14623, USA.
  • Day SW; Department of Biomedical Engineering, Rochester Institute of Technology, Rochester, NY 14623, USA.
Sensors (Basel) ; 24(14)2024 Jul 16.
Article em En | MEDLINE | ID: mdl-39066010
ABSTRACT
Non-invasive monitoring of pulmonary health may be useful for tracking several conditions such as COVID-19 recovery and the progression of pulmonary edema. Some proposed methods use impedance-based technologies to non-invasively measure the thorax impedance as a function of respiration but face challenges that limit the feasibility, accuracy, and practicality of tracking daily changes. In our prior work, we demonstrated a novel approach to monitor respiration by measuring changes in impedance from the back of the thigh. We reported the concept of using thigh-thigh bioimpedance measurements for measuring the respiration rate and demonstrated a linear relationship between the thigh-thigh bioimpedance and lung tidal volume. Here, we investigate the variability in thigh-thigh impedance measurements to further understand the feasibility of the technique for detecting a change in the respiratory status due to disease onset or recovery if used for long-term in-home monitoring. Multiple within-session and day-to-day impedance measurements were collected at 80 kHz using dry electrodes (thigh) and wet electrodes (thorax) across the five healthy subjects, along with simultaneous gold standard spirometer measurements for three consecutive days. The peak-peak bioimpedance measurements were found to be highly correlated (0.94 ± 0.03 for dry electrodes across thigh; 0.92 ± 0.07 for wet electrodes across thorax) with the peak-peak spirometer tidal volume. The data across five subjects indicate that the day-to-day variability in the relationship between impedance and volume for thigh-thigh measurements is smaller (average of 14%) than for the thorax (40%). However, it is affected by food and water and might limit the accuracy of the respiratory tidal volume.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Impedância Elétrica / COVID-19 Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Impedância Elétrica / COVID-19 Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 2024 Tipo de documento: Article