Intraocular pressure measurement and association with corneal biomechanics in patients underwent Descemet's stripping with endothelial keratoplasty: a comparative study.
Chen, Huiyu; Wu, Suqian; Tian, Lijia; Li, Yue; Hong, Jiaxu; Wang, Yulan; Xu, Jianjiang.
Afiliação
  • Chen H; Shanghai Eye Diseases Prevention & Treatment Center/Shanghai Eye Hospital, School of Medicine, Tongji University, Shanghai, China.
  • Wu S; National Clinical Research Center for Eye Diseases, Shanghai Engineering Research Center of Precise Diagnosis and Treatment of Eye Diseases, Shanghai, China.
  • Tian L; Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China.
  • Li Y; NHC Key Laboratory of Myopia and Related Eye Diseases, Chinese Academy of Medical Science, Shanghai, China.
  • Hong J; Shanghai Key Laboratory of Visual Impairment and Restoration, Shanghai, China.
  • Wang Y; Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai, China.
  • Xu J; NHC Key Laboratory of Myopia and Related Eye Diseases, Chinese Academy of Medical Science, Shanghai, China.
Front Med (Lausanne) ; 11: 1384694, 2024.
Article em En | MEDLINE | ID: mdl-39071083
ABSTRACT

Purpose:

To compare corneal biomechanical properties and intraocular pressure (IOP) measurements in patients who underwent Descemet's stripping with endothelial keratoplasty (DSEK) with those of the follow healthy eyes.

Methods:

In this retrospective comparative study, a total of 35 eyes of 35 patients who underwent DSEK by a single surgeon from 2015.02 to 2019.12 were enrolled along with their fellow healthy eyes. Corneal biomechanical parameters were assessed at least 3 months post-DSEK using Corneal Visualization Scheimpflug Technology (CST). IOP was measured by CST, Goldmann applanation tonometry (GAT), and MacKay-Marg tonometer.

Results:

Central corneal thickness (CCT) and stiffness parameter at first applanation (SP-A1) were significantly increased after DSEK when compared to the fellow eyes. In DSEK eyes, biomechanically-corrected intraocular pressure (bIOP) and MacKay-Marg IOP correlated significantly with GAT IOP measurements, with bIOP showed the lowest IOP values. All the IOP values did not correlate with CCT. However, GAT-IOP and MacKay-Marg IOP showed a positive correlation with SP-A1.

Conclusion:

The corneal stiffness increased after DSEK. Central corneal thickness may have less influence than corneal biomechanics on IOP measurements in eyes after DSEK. Biomechanically-corrected IOP obtained by CST seemed to be lower than other tonometry techniques in DSEK eyes, perhaps because of correction for corneal stiffness, CCT and age.