Your browser doesn't support javascript.
loading
Extraction of lumen less pacing leads from the His Bundle and left bundle branch area - outcomes of the high volume center.
Gardas, Rafal; Loboda, Danuta; Biernat, Jolanta; Soral, Tomasz; Kulesza, Piotr; Gladysz-Wanha, Sylwia; Joniec, Michal; Sajdok, Mateusz; Zub, Kamil; Golba, Krzysztof S.
Afiliação
  • Gardas R; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Loboda D; Department of Electrocardiology, Leszek Giec Upper-Silesian Medical Centre of the Silesian Medical University in Katowice.
  • Biernat J; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Soral T; Department of Electrocardiology, Leszek Giec Upper-Silesian Medical Centre of the Silesian Medical University in Katowice.
  • Kulesza P; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Gladysz-Wanha S; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Joniec M; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Sajdok M; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Zub K; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
  • Golba KS; Department of Electrocardiology and Heart Failure, Medical University of Silesia, Katowice, Poland.
Europace ; 2024 Aug 13.
Article em En | MEDLINE | ID: mdl-39137240
ABSTRACT
BACKGROUND AND

AIMS:

The application of conduction system pacing (CSP) in clinical practice is growing, and the need for lead extraction will also increase. The data on outcomes and safety of CSP lead extraction are limited. The aim of this study was to assess procedural outcomes and safety of CSP lead removal.

METHODS:

Forty-seven patients from the EXTRACT Registry with the indication for CSP lead removal were enrolled in the study conducted at the Department of Electrocardiology in Katowice, Poland. Extraction technique, outcomes, safety, and complication were evaluated.

RESULTS:

Forty-three (91.5%) leads were successfully removed, and 41 (87.2%) were removed with traction only. The dwelling time of 28 extracted leads was longer than one year, and the oldest extracted lead was implanted for 89 months. Seven (14.9%) leads were removed from the left bundle branch area and 36 from the His bundle. Transient complete AV block occurred during the procedure in two patients. In 27 out of 31 attempts (87.1%), new CSP leads were implanted nine (33.3%) His bundle pacing leads and 18 (66.7%) left bundle branch area pacing leads.

CONCLUSION:

The CSP lead extraction is safe and feasible with a low complication rate and high rate of CSP lead reimplantation.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article