Your browser doesn't support javascript.
loading
Calculation of response parameters for a neutron long counter instrument.
Ali, Fawaz; Surette, Joel; Atanackovic, Jovica.
Afiliação
  • Ali F; Canadian Nuclear Laboratories, Chalk River, Ontario, Canada, K0J 1J0. Electronic address: fawaz.ali@cnl.ca.
  • Surette J; Canadian Nuclear Laboratories, Chalk River, Ontario, Canada, K0J 1J0.
  • Atanackovic J; Health Physics Laboratory, Ontario Power Generation, Whitby, Ontario, Canada, L1N 9E3.
Appl Radiat Isot ; 214: 111502, 2024 Sep 06.
Article em En | MEDLINE | ID: mdl-39276634
ABSTRACT
The primary goal of this study was to develop a simulation model of a long counter available at Canadian Nuclear Laboratories (CNL). Using the Monte Carlo N-Particle version 6 (MCNP6) code, the model was used to calculate, as a function of incident energy, the number of counts recorded per source neutron, effective centre, and sensitivity. This study also carried out measurements of the neutron emission rate of and direct neutron flux at 2 m from an in-house 252Cf neutron source.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Ano de publicação: 2024 Tipo de documento: Article