Your browser doesn't support javascript.
loading
Trace analysis of the MIF analogue pareptide in blood plasma by high-performance liquid chromatography and short-wavelength excitation fluorometry.
J Chromatogr ; 163(4): 383-9, 1979 Aug 21.
Article em En | MEDLINE | ID: mdl-544604
ABSTRACT
A high-performance liquid chromatographic procedure was developed and applied to analysis of the pharmacologically active MIF analogue pareptide in human plasma. The procedure involves formation of a fluorescent 7-chloro-4-nitrobenzyl-2-oxa-1,3-diazole (NBD-Cl) pareptide derivative followed by separation of the NBD derivative from plasma components on a 30-cm microparticle octadecylsilane bonded column. The separated derivative was quantitated using a short-wavelength excitation fluorometric detector. The detection limit of pareptide in plasma samples was 5 ng or 17 pmoles per ml of plasma. In the absence of plasma, the corresponding on-column detection limit was 0.5 pmoles.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oligopeptídeos / Hormônio Inibidor da Liberação de MSH Limite: Humans Idioma: En Ano de publicação: 1979 Tipo de documento: Article
Buscar no Google