Your browser doesn't support javascript.
loading
A simple and rapid microdiffusion method for blood ammonia using a reflectance meter and a reagent plate, and its clinical evaluation for liver diseases.
Clin Chim Acta ; 144(2-3): 195-202, 1984 Dec 29.
Article em En | MEDLINE | ID: mdl-6529854
ABSTRACT
A simple and rapid procedure using 20 microliters of whole blood is described for estimating blood ammonia by means of a reflectance meter using a reagent plate, which operates by the principle of microdiffusion. The most suitable condition for microdiffusion was at 37 degrees C for 15 min, in which the standard curve was linear, and reproducibility and recovery were sastisfactory. Blood ammonia levels by this method correlate significantly with the values by an ion exchange method in 72 patients with liver disease (r = + 0.970, y = 1.13x - 11). The method has a distinct advantage in sensitivity, simplicity, and rapidity for blood ammonia determination.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Amônia / Hepatopatias Limite: Humans / Male / Middle aged Idioma: En Ano de publicação: 1984 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Amônia / Hepatopatias Limite: Humans / Male / Middle aged Idioma: En Ano de publicação: 1984 Tipo de documento: Article