Your browser doesn't support javascript.
loading
Quantification of pericardial effusions by three-dimensional echocardiography.
Vazquez de Prada, J A; Jiang, L; Handschumacher, M D; Xie, S W; Rivera, J M; Schwammenthal, E; Guerrero, J L; Weyman, A E; Levine, R A; Picard, M H.
Afiliação
  • Vazquez de Prada JA; Cardiac Unit, Massachusetts General Hospital, Boston 02114.
J Am Coll Cardiol ; 24(1): 254-9, 1994 Jul.
Article em En | MEDLINE | ID: mdl-8006275
ABSTRACT

OBJECTIVES:

The purpose of this study was to examine the accuracy of three-dimensional echocardiography for the quantification of asymmetric pericardial effusion volume and to compare this new technique with two-dimensional echocardiography.

BACKGROUND:

Quantification of pericardial effusion by two-dimensional echocardiography relies on a symmetric distribution of the fluid. Three-dimensional echocardiography can quantitate volume without these limitations, but its accuracy for pericardial effusion volume has not yet been assessed.

METHODS:

In six open chest dogs, 41 different asymmetrically distributed pericardial effusions of known volume were created by serial infusions of fluid through a pericardial catheter. The hearts were imaged using an automated echocardiographic method that integrates three-dimensional spatial and imaging data. The surfaces of the pericardial sac and heart were then reconstructed, and the volumes of pericardial effusions were calculated. Two-dimensional echocardiography was performed simultaneously, and volumes were calculated using the prolate ellipsoid method. Asymmetric distribution of the fluid was obtained by applying localized hydrostatic pressure to the pericardium.

RESULTS:

The volumes of pericardial effusion quantified using three-dimensional echocardiography correlated well with actual volumes (y = 1.0x - 1.4, SEE = 7.7 ml, r = 0.98). Two-dimensional echocardiography had an acceptable correlation (y = 1.0x + 2.3, SEE = 23 ml, r = 0.84), but a marked degree of variation from the true value was observed for any individual measurement.

CONCLUSIONS:

Three-dimensional echocardiography accurately quantifies pericardial effusion volume in vivo, even when the fluid is distributed asymmetrically, whereas two-dimensional echocardiography is less reliable. This new technique may be of clinical value in quantitating pericardial effusion, especially in the serial evaluation of asymmetric or loculated effusions.