Your browser doesn't support javascript.
loading
Measurement of primary bremsstrahlung spectrum from an 8-MeV linear accelerator.
Med Phys ; 3(3): 173-5, 1976.
Article em En | MEDLINE | ID: mdl-819770
ABSTRACT
The bremsstrahlung spectrum from an 8-MeV linear accelerator has been measured using a NaI(T1) spectrometer system. The spectrum shows a low-energy cutoff at 0.4 MeV and the maximum photon energy to be approximately 6% greater than the nominal energy. The maximum emission of energy fluence was 1.6 and 1.8 MeV for measured and calculated values, respectively. The fast neutron dose in the photon beam was approximately 0.09% of the x-ray dose. The weighted mean energy was 2.3 MeV, measured value, and 2.4 MeV, calculated value.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radioterapia de Alta Energia Idioma: En Ano de publicação: 1976 Tipo de documento: Article
Buscar no Google