Echo-planar imaging of the liver with a standard MR imaging system.
Radiology
; 189(1): 259-64, 1993 Oct.
Article
em En
| MEDLINE
| ID: mdl-8396785
ABSTRACT
PURPOSE:
A multisection, whole-body echo-planar imaging (EPI) sequence was developed to obtain T2-weighted images of the liver in one 18-second breath hold with a standard magnetic resonance (MR) imaging system. MATERIALS ANDMETHODS:
This capability was achieved by dividing the data acquisition period into eight interleaved segments rather than one or two as implemented previously with EPI systems having high-power gradient subsystems.RESULTS:
The interleaved echo-planar images had excellent depiction of anatomy and no identifiable respiratory artifact. In 26 lesions in 12 patients, the eight-shot echo-planar images (2,000/66 [repetition time msec/echo time msec]) had superior contrast compared with conventional T2-weighted spin-echo (SE) images (2,500/60) by an average factor of 1.22 +/- 0.31 (standard deviation) and an average contrast-to-noise ratio relative to conventional T2-weighted SE images of 0.85 +/- 0.22.CONCLUSION:
With a conventional MR imaging system, breath-hold T2-weighted echo-planar images of the liver are comparable in diagnostic quality to conventional T2-weighted SE images.