Anterior tibial compartment pressures during intermittent sequential pneumatic compression therapy.
Gilbart, M K; Oglivie-Harris, D J; Broadhurst, C; Clarfield, M.
Afiliação
  • Gilbart MK; University of Toronto, Ontario, Canada.
Am J Sports Med ; 23(6): 769-72, 1995.
Article em En | MEDLINE | ID: mdl-8600749
ABSTRACT
We studied the anterior tibial compartment pressures during the application of a JOBST sequential intermittent pneumatic compression device on 5 healthy human volunteers (10 legs). Intracompartmental pressures were measured using a slit catheter. The measurements of interstitial pressures were highest at maximal calf inflation, and pressures were increased for approximately 120 seconds during each cycle. Pressure measurements in the inflated pressure sleeve varied less than 10% with the measured anterior tibial compartment pressures during intermittent pneumatic compression therapy. This intermittent pneumatic compression device may elevate intramuscular pressure significantly above that necessary to render muscle ischemic. However, these periods of pressure elevation are not long enough to produce any significant adverse effects, and the beneficial effects of decreased edema fluid may be safely realized.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tíbia / Músculo Esquelético / Trajes Gravitacionais Tipo de estudo: Etiology_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 1995 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tíbia / Músculo Esquelético / Trajes Gravitacionais Tipo de estudo: Etiology_studies Limite: Adult / Female / Humans / Male Idioma: En Ano de publicação: 1995 Tipo de documento: Article