Stimulating technetium-99m cerebral perfusion studies with a three-dimensional Hoffmann brain phantom: collimator and filter selection in SPECT neuroimaging.
Ann Nucl Med
; 10(1): 153-60, 1996 Feb.
Article
em En
| MEDLINE
| ID: mdl-8814722
The choice of collimator and the selection of a filter can affect the quality of clinical SPECT images of the brain. The compromises that 4 different collimators make between spatial resolution and sensitivity were studied by imaging a three-dimensional Hoffmann brain phantom. The planar data were acquired with each collimator on a three-headed SPECT system and were reconstructed with both a standard Butterworth filter and a Wiener pre-filter. The reconstructed images were then evaluated by specialists in nuclear medicine and were also quantitatively analyzed with specific regions of interest (ROI) in the brain. All observers preferred the Wiener filter reconstructed images regardless of the collimator used to acquire the planar images. With this filter, the ultrahigh-resolution fan-beam collimator was the most subjectively preferable and quantitatively produced the highest contrast ratios. The findings support suggestions that higher resolution collimators are preferable to higher sensitivity collimators, and indicate that fan-beam collimators are preferable to parallel-hole collimators for clinical SPECT studies of cerebral perfusion. The results also suggest that Wiener filter enhances the quality of SPECT brain images regardless of which collimator is used to acquire the data.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Encéfalo
/
Tomografia Computadorizada de Emissão de Fóton Único
/
Tecnécio
Tipo de estudo:
Diagnostic_studies
/
Evaluation_studies
/
Prognostic_studies
Limite:
Humans
Idioma:
En
Ano de publicação:
1996
Tipo de documento:
Article