Spurious detection of ventricular fibrillation by a transvenous implantable cardioverter defibrillator during percutaneous transluminal coronary angioplasty.
Pacing Clin Electrophysiol
; 20(1 Pt 1): 136-7, 1997 Jan.
Article
em En
| MEDLINE
| ID: mdl-9121962
ABSTRACT
This case describes another example of inappropriate implantable defibrillator shock. We also show that this phenomenon resulted from electromagnetic current traveling down a coronary wire during coronary angioplasty.