Cardiac pacemaker dysfunction in children after thoracic drainage catheter manipulation.
Lobdell, K W; Walters, H L; Hudson, C; Hakimi, M.
Afiliação
  • Lobdell KW; Department of Cardiovascular Surgery, Children's Hospital of Michigan, Detroit 48201, USA.
Ann Thorac Surg ; 63(5): 1466-7, 1997 May.
Article em En | MEDLINE | ID: mdl-9146348
ABSTRACT
Two children underwent placement of permanent, epicardial-lead, dual-chamber, unipolar pacemaker systems for complete heart block. Postoperatively, both patients demonstrated subcutaneous emphysema-in the area of their pulse generators-temporally related to thoracic catheter manipulation. Acutely, each situation was managed with manual compression of the pulse generator, ascertaining appropriate pacemaker sensing and pacing. Maintenance of compression with pressure dressings, vigilant observation/monitoring, and education of the care givers resulted in satisfactory pacemaker function without invasive intervention.