Your browser doesn't support javascript.
loading
Defibrillation efficacy of different electrode placements in a human thorax model.
de Jongh, A L; Entcheva, E G; Replogle, J A; Booker, R S; Kenknight, B H; Claydon, F J.
Afiliação
  • de Jongh AL; Department of Biomedical Engineering, University of Memphis, Tennessee 38152-6582, USA.
Pacing Clin Electrophysiol ; 22(1 Pt 2): 152-7, 1999 Jan.
Article em En | MEDLINE | ID: mdl-9990621
ABSTRACT
The objective of this study was to measure the defibrillation threshold (DFT) associated with different electrode placements using a three-dimensional anatomically realistic finite element model of the human thorax. Coil electrodes (Endotak DSP, model 125, Guidant/CPI) were placed in the RV apex along the lateral wall (RV), withdrawn 10 mm away from the RV apex along the lateral wall (RVprox), in the RV apex along the anterior septum (RVseptal), and in the SVC. An active pulse generator (can) was placed in the subcutaneous prepectoral space. Five electrode configurations were studied RV-->SVC, RVprox-->SVC, RVSEPTAL-->SVC, RV-->Can, and RV-->SVC + Can. DFTs are defined as the energy required to produce a potential gradient of at least 5 V/cm in 95% of the ventricular myocardium. DFTs for RV-->SVC, RVprox-->SVC, RVseptal-->SVC, RV-->Can, and RV-->SVC + Can were 10, 16, 7, 9, and 6 J, respectively. The DFTs measured at each configuration fell within one standard deviation of the mean DFTs reported in clinical studies using the Endotak leads. The relative changes in DFT among electrode configurations also compared favorably. This computer model allows measurements of DFT or other defibrillation parameters with several different electrode configurations saving time and cost of clinical studies.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tórax / Simulação por Computador / Cardioversão Elétrica / Desfibriladores Implantáveis / Modelos Cardiovasculares Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans / Male Idioma: En Ano de publicação: 1999 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tórax / Simulação por Computador / Cardioversão Elétrica / Desfibriladores Implantáveis / Modelos Cardiovasculares Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans / Male Idioma: En Ano de publicação: 1999 Tipo de documento: Article