Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros








Base de dados
Intervalo de ano de publicação
1.
Magn Reson Med ; 2024 Jul 12.
Artigo em Inglês | MEDLINE | ID: mdl-38997798

RESUMO

PURPOSE: Breath-held fat-suppressed volumetric T1-weighted MRI is an important and widely-used technique for evaluating the abdomen. Both fat-saturation and Dixon-based fat-suppression methods are used at conventional field strengths; however, both have challenges at lower field strengths (<1.5T) due to insufficient fat suppression and/or inadequate resolution. Specifically, at lower field strengths, fat saturation often fails due to the short T1 of lipid; and Cartesian Dixon imaging provides poor spatial resolution due to the need for a long ∆TE, due to the smaller ∆f between water and lipid. The purpose of this work is to demonstrate a new approach capable of simultaneously achieving excellent fat suppression and high spatial resolution on a 0.55T whole-body system. METHODS: We applied 3D stack-of-spirals Dixon imaging at 0.55T, with compensation of concomitant field phase during reconstruction. The spiral readouts make efficient use of the requisite ∆TE. We compared this with 3D Cartesian Dixon imaging. Experiments were performed in 2 healthy and 10 elevated liver fat volunteers. RESULTS: Stack-of-spirals Dixon imaging at 0.55T makes excellent use of the required ∆TE, provided high SNR efficiency and finer spatial resolution (1.7 × 1.7 × 5 mm3) compared Cartesian Dixon (3.5 × 3.5 × 5 mm3), within a 17-s breath-hold. We observed successful fat suppression, and improved definition of structures such as the liver, kidneys, and bowel. CONCLUSION: We demonstrate that high-resolution single breath-hold volumetric abdominal T1-weighted imaging is feasible at 0.55T using spiral sampling and concomitant field correction. This is an attractive alternative to existing Cartesian-based methods, as it simultaneously provides high-resolution and excellent fat-suppression.

2.
Magn Reson Med ; 92(4): 1649-1657, 2024 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-38725132

RESUMO

PURPOSE: To investigate the feasibility of diffusion tensor brain imaging at 0.55T with comparisons against 3T. METHODS: Diffusion tensor imaging data with 2 mm isotropic resolution was acquired on a cohort of five healthy subjects using both 0.55T and 3T scanners. The signal-to-noise ratio (SNR) of the 0.55T data was improved using a previous SNR-enhancing joint reconstruction method that jointly reconstructs the entire set of diffusion weighted images from k-space using shared-edge constraints. Quantitative diffusion tensor parameters were estimated and compared across field strengths. We also performed a test-retest assessment of repeatability at each field strength. RESULTS: After applying SNR-enhancing joint reconstruction, the diffusion tensor parameters obtained from 0.55T data were strongly correlated ( R 2 ≥ 0 . 70 $$ {R}^2\ge 0.70 $$ ) with those obtained from 3T data. Test-retest analysis showed that SNR-enhancing reconstruction improved the repeatability of the 0.55T diffusion tensor parameters. CONCLUSION: High-resolution in vivo diffusion MRI of the human brain is feasible at 0.55T when appropriate noise-mitigation strategies are applied.


Assuntos
Encéfalo , Imagem de Tensor de Difusão , Estudos de Viabilidade , Processamento de Imagem Assistida por Computador , Razão Sinal-Ruído , Humanos , Encéfalo/diagnóstico por imagem , Imagem de Tensor de Difusão/métodos , Masculino , Adulto , Reprodutibilidade dos Testes , Feminino , Processamento de Imagem Assistida por Computador/métodos , Algoritmos , Voluntários Saudáveis
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA