1.
Phys Med ; 111: 102617, 2023 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-37290226

RESUMO

PURPOSE: This work aimed to validate Monte Carlo (MC) simulated cardiac phantoms for the evaluation of planar- and SPECT-gated-blood-pool (GBP-P and GBP-S) studies. METHODS: A comparison of gamma camera system performance criteria measurements (energy resolution, spatial resolution, sensitivity) with MC simulations was conducted. Furthermore, the accuracy of measured and simulated volumes of two stereolithography-printed cardiac phantoms (based on 4D-XCAT phantoms) was assessed. Finally, the simulated GBP-P and GBP-S XCAT studies were validated by comparing calculated left ventricular ejection fraction (LVEF) and ventricle volume values with known parameters. RESULTS: The simulated performance criteria compared well with measured values (energy resolution difference: 0.1 ± 0.10%; spatial resolution (full width at half maximum) difference ≤ 0.5 ± 0.8 mm and system sensitivity difference ≤ 6.2 ± 0.62cps/MBq). The measured and simulated cardiac phantoms were in good agreement; the left anterior oblique views compared well. This is supported by line profiles through these phantoms and on average, simulated counts were 5.8% lower than measured counts. The LVEF values calculated from the GBP-P and GBP-S simulated data differ from known values (2.8 ± 0.64% and 0.8 ± 0.52%). The differences between the known XCAT LV volumes and simulated GBP-S calculated volumes were -1.2 ± 1.91 ml and -1.5 ± 0.96 ml for the end-diastolic and end-systolic volumes. CONCLUSION: The MC-simulated cardiac phantom has been validated successfully. Stereolithography-printing allows researchers to create clinically realistic organ phantoms and is a valuable tool for validating MC simulations and clinical software. By conducting GBP simulation studies with various XCAT models, the user will be able to generate GBP-P and GBP-S databases for future software evaluation.


Assuntos
Coração , Função Ventricular Esquerda , Volume Sistólico , Coração/diagnóstico por imagem , Tomografia Computadorizada de Emissão de Fóton Único/métodos , Software , Imagens de Fantasmas
2.
Heliyon ; 8(7): e09830, 2022 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-35865988

RESUMO

Purpose: Virtual dosimetry using voxel-based patient-specific phantoms and Monte Carlo (MC) simulations offer the advantage of having a gold standard against which absorbed doses may be benchmarked to establish the dosimetry accuracy. Furthermore, these reference values assist in investigating the accuracy of the absorbed dose methodologies from different software programs. Therefore, this study aimed to compare the accuracy of the absorbed doses computed using LundADose and OLINDA/EXM 1.0. Methods: The accuracy was based on 177Lu-DOTATATE distributions of three voxel-based phantoms. SPECT projection images were simulated for 1, 24, 96, and 168 h post-administration and reconstructed with LundADose using 3D OS-EM reconstruction. Mono-exponential curves were fitted to the bio-kinetic data for the kidneys, liver, spleen, and tumours resulting in SPECT time-integrated activity (SPECT-TIA). The SPECT-TIA were used to compute mean absorbed doses using LundADose (LND-DSPECT) and OLINDA (OLINDA-DSPECT) for the organs. Pre-defined true activity images, were used to obtain TRUE-TIA and, together with full MC simulations, computed the true doses (MC-DTrue). The dosimetry accuracy was assessed by comparing LND-DSPECT and OLINDA-DSPECT to MC-DTrue. Results: Overall, the results presented an overestimation of the mean absorbed dose by LND-DSPECT compared to the MC-DTrue with a dosimetry accuracy ≤6.6%. This was attributed to spill-out activity from the reconstructed LND-DSPECT, resulting in a higher dose contribution than the MC-DTrue. There was a general underestimation (<8.1%) of OLINDA-DSPECT compared to MC-DTrue attributed to the geometrical difference in shape between the voxel-based phantoms and the OLINDA models. Furthermore, OLINDA-DSPECT considers self-doses while MC-DTrue reflects self-doses plus cross-doses. Conclusion: The better than 10% accuracy suggests that the mean dose values obtained with LND-DSPECT and OLINDA-DSPECT approximate the true values. The mean absorbed doses of the two software programs, and the gold standard were comparable. This work shall be of use for optimising 177Lu dosimetry for clinical applications.

3.
EJNMMI Phys ; 8(1): 27, 2021 Mar 18.
Artigo em Inglês | MEDLINE | ID: mdl-33738605

RESUMO

BACKGROUND: Different gamma camera calibration factor (CF) geometries have been proposed to convert SPECT data into units of activity concentration. However, no consensus has been reached on a standardised geometry. The CF is dependent on the selected geometry and is further affected by partial volume effects. This study investigated the effect of two CF geometries and their corresponding recovery coefficients (RCs) on the quantification accuracy of 177Lu SPECT images using Monte Carlo simulations. METHODS: The CF geometries investigated were (i) a radioactive-sphere surrounded by non-radioactive water (sphere-CF) and (ii) a cylindrical phantom uniformly filled with radioactive water (cylinder-CF). Recovery coefficients were obtained using the sphere-CF and cylinder-CF, yielding the sphere-RC and cylinder-RC values, respectively, for partial volume correction (PVC). The quantification accuracy was evaluated using four different-sized spheres (15.6-65.4 ml) and a kidney model with known activity concentrations inside a cylindrical, torso and patient phantom. Images were reconstructed with the 3D OS-EM algorithm incorporating attenuation, scatter and detector-response corrections. Segmentation was performed using the physical size and a small cylindrical volume inside the cylinder for the sphere-CF and cylinder-CF, respectively. RESULTS: The sphere quantification error (without PVC) was better for the sphere-CF (≤ - 5.54%) compared to the cylinder-CF (≤ - 20.90%), attributed to the similar geometry of the quantified and CF spheres. Partial volume correction yielded comparable results for the sphere-CF-RC (≤ 3.47%) and cylinder-CF-RC (≤ 3.53%). The accuracy of the kidney quantification was poorer (≤ 22.34%) for the sphere-CF without PVC compared to the cylinder-CF (≤ 2.44%). With PVC, the kidney quantification results improved and compared well for the sphere-CF-RC (≤ 3.50%) and the cylinder-CF-RC (≤ 3.45%). CONCLUSION: The study demonstrated that upon careful selection of CF-RC combinations, comparable quantification errors (≤ 3.53%) were obtained between the sphere-CF-RC and cylinder-CF-RC, when all corrections were applied.