1.
Magn Reson Med ; 83(1): 352-366, 2020 01.
Artículo en Inglés | MEDLINE | ID: mdl-31385628

RESUMEN

PURPOSE: To establish peripheral nerve stimulation (PNS) thresholds for an ultra-high performance magnetic field gradient subsystem (simultaneous 200-mT/m gradient amplitude and 500-T/m/s gradient slew rate; 1 MVA per axis [MAGNUS]) designed for neuroimaging with asymmetric transverse gradients and 42-cm inner diameter, and to determine PNS threshold dependencies on gender, age, patient positioning within the gradient subsystem, and anatomical landmarks. METHODS: The MAGNUS head gradient was installed in a whole-body 3T scanner with a custom 16-rung bird-cage transmit/receive RF coil compatible with phased-array receiver brain coils. Twenty adult subjects (10 male, mean ± SD age = 40.4 ± 11.1 years) underwent the imaging and PNS study. The tests were repeated by displacing subject positions by 2-4 cm in the superior-inferior and anterior-posterior directions. RESULTS: The x-axis (left-right) yielded mostly facial stimulation, with mean ΔGmin = 111 ± 6 mT/m, chronaxie = 766 ± 76 µsec. The z-axis (superior-inferior) yielded mostly chest/shoulder stimulation (123 ± 7 mT/m, 620 ± 62 µsec). Y-axis (anterior-posterior) stimulation was negligible. X-axis and z-axis thresholds tended to increase with age, and there was negligible dependency with gender. Translation in the inferior and posterior directions tended to increase the x-axis and z-axis thresholds, respectively. Electric field simulations showed good agreement with the PNS results. Imaging at MAGNUS gradient performance with increased PNS threshold provided a 35% reduction in noise-to-diffusion contrast as compared with whole-body performance (80 mT/m gradient amplitude, 200 T/m/sec gradient slew rate). CONCLUSION: The PNS threshold of MAGNUS is significantly higher than that for whole-body gradients, which allows for diffusion gradients with short rise times (under 1 msec), important for interrogating brain microstructure length scales.


Asunto(s)
Encéfalo/diagnóstico por imagen , Estimulación Eléctrica , Campos Magnéticos , Neuroimagen/instrumentación , Neuroimagen/métodos , Nervios Periféricos/diagnóstico por imagen , Sistema Nervioso Periférico/diagnóstico por imagen , Adulto , Algoritmos , Diseño de Equipo , Femenino , Humanos , Procesamiento de Imagen Asistido por Computador , Imagen por Resonancia Magnética , Masculino , Persona de Mediana Edad , Nervios Periféricos/fisiología , Fantasmas de Imagen , Reproducibilidad de los Resultados , Imagen de Cuerpo Entero
2.
Magn Reson Med ; 83(6): 2356-2369, 2020 06.
Artículo en Inglés | MEDLINE | ID: mdl-31763726

RESUMEN

PURPOSE: To develop a highly efficient magnetic field gradient coil for head imaging that achieves 200 mT/m and 500 T/m/s on each axis using a standard 1 MVA gradient driver in clinical whole-body 3.0T MR magnet. METHODS: A 42-cm inner diameter head-gradient used the available 89- to 91-cm warm bore space in a whole-body 3.0T magnet by increasing the radial separation between the primary and the shield coil windings to 18.6 cm. This required the removal of the standard whole-body gradient and radiofrequency coils. To achieve a coil efficiency ~4× that of whole-body gradients, a double-layer primary coil design with asymmetric x-y axes, and symmetric z-axis was used. The use of all-hollow conductor with direct fluid cooling of the gradient coil enabled ≥50 kW of total heat dissipation. RESULTS: This design achieved a coil efficiency of 0.32 mT/m/A, allowing 200 mT/m and 500 T/m/s for a 620 A/1500 V driver. The gradient coil yielded substantially reduced echo spacing, and minimum repetition time and echo time. In high b = 10,000 s/mm2 diffusion, echo time (TE) < 50 ms was achieved (>50% reduction compared with whole-body gradients). The gradient coil passed the American College of Radiology tests for gradient linearity and distortion, and met acoustic requirements for nonsignificant risk operation. CONCLUSIONS: Ultra-high gradient coil performance was achieved for head imaging without substantial increases in gradient driver power in a whole-body 3.0T magnet after removing the standard gradient coil. As such, any clinical whole-body 3.0T MR system could be upgraded with 3-4× improvement in gradient performance for brain imaging.


Asunto(s)
Encéfalo , Imagen por Resonancia Magnética , Acústica , Encéfalo/diagnóstico por imagen , Diseño de Equipo , Cabeza/diagnóstico por imagen , Humanos , Campos Magnéticos
3.
Magn Reson Med ; 80(5): 2232-2245, 2018 11.
Artículo en Inglés | MEDLINE | ID: mdl-29536587

RESUMEN

PURPOSE: To build and evaluate a small-footprint, lightweight, high-performance 3T MRI scanner for advanced brain imaging with image quality that is equal to or better than conventional whole-body clinical 3T MRI scanners, while achieving substantial reductions in installation costs. METHODS: A conduction-cooled magnet was developed that uses less than 12 liters of liquid helium in a gas-charged sealed system, and standard NbTi wire, and weighs approximately 2000 kg. A 42-cm inner-diameter gradient coil with asymmetric transverse axes was developed to provide patient access for head and extremity exams, while minimizing magnet-gradient interactions that adversely affect image quality. The gradient coil was designed to achieve simultaneous operation of 80-mT/m peak gradient amplitude at a slew rate of 700 T/m/s on each gradient axis using readily available 1-MVA gradient drivers. RESULTS: In a comparison of anatomical imaging in 16 patients using T2 -weighted 3D fluid-attenuated inversion recovery (FLAIR) between the compact 3T and whole-body 3T, image quality was assessed as equivalent to or better across several metrics. The ability to fully use a high slew rate of 700 T/m/s simultaneously with 80-mT/m maximum gradient amplitude resulted in improvements in image quality across EPI, DWI, and anatomical imaging of the brain. CONCLUSIONS: The compact 3T MRI system has been in continuous operation at the Mayo Clinic since March 2016. To date, over 200 patient studies have been completed, including 96 comparison studies with a clinical 3T whole-body MRI. The increased gradient performance has reliably resulted in consistently improved image quality.


Asunto(s)
Imagen por Resonancia Magnética/instrumentación , Imagen de Cuerpo Entero/instrumentación , Encéfalo/diagnóstico por imagen , Diseño de Equipo , Femenino , Humanos , Imagenología Tridimensional , Imanes , Masculino , Fantasmas de Imagen , Relación Señal-Ruido
4.
Magn Reson Med ; 76(6): 1939-1950, 2016 12.
Artículo en Inglés | MEDLINE | ID: mdl-26628078

RESUMEN

PURPOSE: To characterize peripheral nerve stimulation (PNS) of an asymmetric head-only gradient coil that is compatible with a commercial high-channel-count receive-only array. METHODS: Two prototypes of an asymmetric head-only gradient coil set with a 42-cm inner diameter were constructed for brain imaging at 3T with maximum performance specifications of up to 85 mT/m and 708 T/m/s. Tests were performed in 24 volunteers to measure PNS thresholds with the transverse (x = left-right; y = anterior-posterior [A/P]) gradient coils of both prototypes. Fourteen of these 24 volunteers were also tested for the z-gradient PNS in the second prototype and were scanned with high-slew-rate echo planar imaging (EPI) immediately after the PNS tests. RESULTS: For both prototypes, the y-gradient PNS threshold was markedly higher than the x-gradient threshold. The z-gradient threshold was intermediate between those for the x- and y-coils. Of the 24 volunteers, only two experienced y-gradient PNS at 80 mT/m and 500 T/m/s. All volunteers underwent the EPI scan without PNS when the readout direction was set to A/P. CONCLUSION: Measured PNS characteristics of asymmetric head-only gradient coil prototypes indicate that such coils, especially in the A/P direction, can be used for fast EPI readout in high-performance neuroimaging scans with substantially reduced PNS concerns compared with conventional whole body gradient coils. Magn Reson Med 76:1939-1950, 2016. © 2015 International Society for Magnetic Resonance in Medicine.


Asunto(s)
Encéfalo/fisiología , Terapia por Estimulación Eléctrica/instrumentación , Neuroestimuladores Implantables , Imagen por Resonancia Magnética/instrumentación , Nervios Periféricos/fisiología , Encéfalo/anatomía & histología , Terapia por Estimulación Eléctrica/métodos , Diseño de Equipo , Análisis de Falla de Equipo , Humanos , Fantasmas de Imagen , Reproducibilidad de los Resultados , Sensibilidad y Especificidad
5.
J Magn Reson Imaging ; 44(3): 653-64, 2016 09.
Artículo en Inglés | MEDLINE | ID: mdl-26921117

RESUMEN

PURPOSE: To investigate the effects on echo planar imaging (EPI) distortion of using high gradient slew rates (SR) of up to 700 T/m/s for in vivo human brain imaging, with a dedicated, head-only gradient coil. MATERIALS AND METHODS: Simulation studies were first performed to determine the expected echo spacing and distortion reduction in EPI. A head gradient of 42-cm inner diameter and with asymmetric transverse coils was then installed in a whole-body, conventional 3T magnetic resonance imaging (MRI) system. Human subject imaging was performed on five subjects to determine the effects of EPI on echo spacing and signal dropout at various gradient slew rates. The feasibility of whole-brain imaging at 1.5 mm-isotropic spatial resolution was demonstrated with gradient-echo and spin-echo diffusion-weighted EPI. RESULTS: As compared to a whole-body gradient coil, the EPI echo spacing in the head-only gradient coil was reduced by 48%. Simulation and in vivo results, respectively, showed up to 25-26% and 19% improvement in signal dropout. Whole-brain imaging with EPI at 1.5 mm spatial resolution provided good whole-brain coverage, spatial linearity, and low spatial distortion effects. CONCLUSION: Our results of human brain imaging with EPI using the compact head gradient coil at slew rates higher than in conventional whole-body MR systems demonstrate substantially improved image distortion, and point to a potential for benefits to non-EPI pulse sequences. J. Magn. Reson. Imaging 2016;44:653-664.


Asunto(s)
Artefactos , Encéfalo/anatomía & histología , Encéfalo/diagnóstico por imagen , Imagen Eco-Planar/instrumentación , Imagen Eco-Planar/métodos , Aumento de la Imagen/instrumentación , Aumento de la Imagen/métodos , Diseño de Equipo , Análisis de Falla de Equipo , Humanos , Interpretación de Imagen Asistida por Computador/instrumentación , Interpretación de Imagen Asistida por Computador/métodos , Proyectos Piloto , Reproducibilidad de los Resultados , Sensibilidad y Especificidad , Transductores
6.
NMR Biomed ; 22(8): 867-73, 2009 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-19489035

RESUMEN

The ability to accelerate the spatial encoding process during a chemical shift imaging (CSI) scan of hyperpolarized compounds is demonstrated through parallel imaging. A hardware setup designed to simultaneously acquire (13)C data from multiple receivers is presented here. A system consisting of four preamplifiers, four gain stages, a transmit coil, and a four receive channel rat coil was built for single channel excitation and simultaneous multi-channel detection of (13)C signals. The hardware setup was integrated with commercial scanner electronics, allowing the system to function similar to a conventional proton multi-channel setup, except at a different frequency. The ability to perform parallel imaging is demonstrated in vivo. CSI data from the accelerated scans are reconstructed using a self-calibrated multi-spectral parallel imaging algorithm, by using lower resolution coil sensitivity maps obtained from the central region of k-space. The advantages and disadvantages of parallel imaging in the context of imaging hyperpolarized compounds are discussed.


Asunto(s)
Radioisótopos de Carbono/química , Aumento de la Imagen , Imagen por Resonancia Magnética , Ácido Pirúvico/química , Algoritmos , Animales , Humanos , Aumento de la Imagen/instrumentación , Aumento de la Imagen/métodos , Imagen por Resonancia Magnética/instrumentación , Imagen por Resonancia Magnética/métodos , Ratas , Ratas Sprague-Dawley
7.
Magn Reson Med ; 60(4): 928-33, 2008 Oct.
Artículo en Inglés | MEDLINE | ID: mdl-18816813

RESUMEN

Imaging exams involving hyperpolarized, (13)C-labeled compounds require novel RF coils for efficient signal utilization. While (13)C coils are required for mapping the spatial distribution of the hyperpolarized compounds, imaging/pulsing at different frequencies is also needed for scan setup steps prior to the image acquisition. Imaging/pulsing at the (1)H frequency is typically used for anatomical localization and shimming. Flip angle (FA) calibration, which is difficult or impossible to achieve at the (13)C frequency, can be accurately performed at the (23)Na frequency using the natural abundance signal that exists in any living tissue. We demonstrate here a single RF resonant structure that is capable of operating linearly at the (1)H and (23)Na frequencies for scan setup steps, and in quadrature at the (13)C frequency for imaging. Images at the three resonant frequencies of this coil are presented from an exam involving hyperpolarized (13)C compounds in vivo.


Asunto(s)
Radioisótopos de Carbono/análisis , Aumento de la Imagen/instrumentación , Imagen por Resonancia Magnética/instrumentación , Espectroscopía de Resonancia Magnética/instrumentación , Transductores , Animales , Diseño de Equipo , Análisis de Falla de Equipo , Fantasmas de Imagen , Protones , Ondas de Radio , Ratas , Ratas Sprague-Dawley , Reproducibilidad de los Resultados , Sensibilidad y Especificidad
8.
Echocardiography ; 13(6): 677-684, 1996 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-11442987

RESUMEN

This work extends the clinical benefits of phased array transesophageal echocardiography with high detail and contrast resolution to include neonatal patients. We have built several prototype, 64-element, 7.5-MHz phased array transducers housed in 6.2-mm endoscope shafts for use with commercially available imaging systems. The acoustic design is standard, but the miniaturized packaging of the electrical connections was quite challenging. The endoscopes demonstrate very good structural resolution and excellent sensitivity for color flow imaging and continuous-wave Doppler. They have been used on patients as small as 1.9 kg, frequently as an anatomical guide during catheter-based interventions and during congenital heart surgery. Array test data and representative clinical studies are shown. (ECHOCARDIOGRAPHY, Volume 13, November 1996)

9.
J Magn Reson Imaging ; 28(5): 1219-25, 2008 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-18972330

RESUMEN

PURPOSE: To determine whether the promise of high-density many-coil MRI receiver arrays for enabling highly accelerated parallel imaging can be realized in practice. MATERIALS AND METHODS: A 128-channel body receiver-coil array and custom MRI system were developed. The array comprises two clamshells containing 64 coils each, with the posterior array built to maximize signal-to-noise ratio (SNR) and the anterior array design incorporating considerations of weight and flexibility as well. Phantom imaging and human body imaging were performed using a variety of reduction factors and 2D and 3D pulse sequences. RESULTS: The ratio of SNR relative to a 32-element array of similar footprint was 1.03 in the center of an elliptical loading phantom and 1.7 on average in the outer regions. Maximum g-factors dropped from 5.5 (for 32 channels) to 2.0 (for 128 channels) for 4x4 acceleration and from 25 to 3.3 for 5x5 acceleration. Residual aliasing artifacts for a right/left (R/L) reduction factor of 8 in human body imaging were significantly reduced relative to the 32-channel array. CONCLUSION: MRI with a large number of receiver channels enables significantly higher acceleration factors for parallel imaging and improved SNR, provided losses from the coils and electronics are kept negligible.


Asunto(s)
Aumento de la Imagen/instrumentación , Imagen por Resonancia Magnética/instrumentación , Magnetismo/instrumentación , Transductores , Imagen de Cuerpo Entero/instrumentación , Diseño de Equipo , Análisis de Falla de Equipo , Humanos , Fantasmas de Imagen , Reproducibilidad de los Resultados , Sensibilidad y Especificidad