Radiat Prot Dosimetry ; 168(1): 40-5, 2016 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-25737582

RESUMEN

The response of LiF:Mg,Cu,P thermoluminescence dosemeters (type GR-200A) to monoenergetic radiation of energy 28, 35, 38 and 40 keV was evaluated with respect to irradiation with a calibrated (60)Co gamma-ray source. High-precision measurements of the relative air kerma response performed at the SYRMEP beamline of the ELETTRA synchrotron radiation facility (Trieste, Italy) showed a significant deviation of the average response to low-energy X-rays from that to (60)Co, with an over-response from 6 % (at 28 keV) to 22 % (at 40 keV). These data are not consistent with literature data for these dosemeters, where model predictions gave deviation from unity of the relative air kerma response of about 10 %. The authors conclude for the need of additional determinations of the low-energy relative response of GR-200A dosemeters, covering a wider range of monoenergetic energies sampled at a fine energy step, as planned in future experiments by their group at the ELETTRA facility.


Asunto(s)
Sincrotrones , Dosimetría Termoluminiscente/métodos , Aire , Calibración , Radioisótopos de Cobalto/análisis , Cobre/química , Fluoruros/química , Rayos gamma , Compuestos de Litio/química , Magnesio/química , Fósforo/química , Fotones , Dosis de Radiación , Rayos X