Article in Korean | WPRIM | ID: wpr-56832

ABSTRACT

Heated humidifiers are used to prevent respiratory complications associated with mechanical ventilation with an endotracheal tube. One of the heated humidifiers uses an air dryer filter in order to remove humidity from air before it enters the anesthetic machine. We report a case in which the air dryer filter was unintentionally misconnected to the inspiratory limb of the circuit, which caused insufficient volatile anesthetic concentration during anesthesia induction.


Subject(s)
Adsorption , Anesthesia , Extremities , Hot Temperature , Humidity , Isoflurane , Respiration, Artificial , Silica Gel