Dabigatran en plasma: tiempos de trombina diluido, Ecarin y veneno de víbora de Russell diluido / Plasma dabigatran: diluted thrombin time, Ecarin and diluted Russell Viper Venom Time / Dabigatrana em plasma: tempos de trombina diluído, Ecarin e veneno de víbora de Russell diluído
Martinuzzo, Marta; Duboscq, Cristina; López, Marina Sol; Montenegro, Marina; Viñuales, Susana; Girardi, Beatriz; Penchasky, Diana; Otero, Victoria; Ceresetto, José; Stemmelin, German; Barrera, Luis; Otaso, Juan Carlos; Oyhamburu, José.
  • Martinuzzo, Marta; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Duboscq, Cristina; Hospital Británico de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • López, Marina Sol; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Montenegro, Marina; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Viñuales, Susana; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Girardi, Beatriz; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Penchasky, Diana; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Otero, Victoria; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Ceresetto, José; Hospital Británico de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Stemmelin, German; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Barrera, Luis; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Otaso, Juan Carlos; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
  • Oyhamburu, José; Hospital Italiano de Buenos Aires. Ciudad Autónoma de Buenos Aires. AR
Acta bioquím. clín. latinoam ; 50(2): 193-203, jun. 2016. graf, tab
Article in Spanish | LILACS | ID: biblio-837598
RESUMEN
Los objetivos del trabajo fueron verificar la calidad analítica del ensayo tiempo de trombina diluido (DTI) para medición de la concentración plasmática (cc) de dabigatran comparando dos coagulómetros de detección foto-óptica, comparar los resultados con el tiempo de Ecarin (ECT) y correlacionar las cc con las pruebas básicas de coagulación Tiempo de protrombina (TP), APTT y Tiempo de trombina (TT), y tiempo de veneno de víbora de Russell con fosfolípidos concentrados (DRVVTC). Se tomaron 43 muestras de plasma en el valle (10-14 h de la última toma) de 40 pacientes que recibían dabigatran. DTI y ECT presentaron (%) repetitividad <5,4% y <7,5%, CV interensayo <6% y <9%, respectivamente, en el protocolo EP15A2, aceptables para un Error Total permitido (TEa) <15%. Las cc medidas en pacientes fueron mediana 83 (4-945) ng/mL. La comparación de equipos ACL TOP 300 y 500 dio resultados equivalentes por procedimiento alternativo de comparación de métodos. La comparación ECT vs. DTI fue satisfactoria por regresión de Deming (pendiente 1,143, ordenada al origen -19,33). Las correlaciones de cc vs. APTT, TP y DRVVTC fueron moderadas y no lineales tendiendo a plateau a cc>350 ng/mL, r2 0,59, 0,66 y 0,59, respectivamente. El TT fue extremadamente sensible >120 s a cc 50 ng/mL. DTI presentó un buen desempeño analítico y permitió cuantificar dabigatran plasmático a cc bajas y altas en ambos equipos utilizados. ECT presentó resultados comparables a DTI. Se verifica una correlación moderada entre cc de dabigatran y las pruebas clásicas y DRVVTC, pudiendo ser estimadores de cc a partir de los 50 ng/mL.
ABSTRACT
The aims of the study were to verify the analytical performance of Dilute Thrombin Time (DTI) test to measure plasma dabigatran concentration (cc) in two photo-optical coagulometers, compare Ecarin clotting Time (ECT) and DTI results, and correlate cc with classical coagulation tests prothrombin time (PT), APTT, thrombin time (TT) and diluted Russell Viper Venom Time tests with high phospholipid concentration (DRVVTC). Forty three plasma samples from 40 patients taking dabigatran were drown at through (10-14 hs.since last dose). DTI and ECT showed repetitivity (%) <5.4% and <7.5%, interassay CV <6% and <9%, respectively, following EP15A2 protocol, acceptable considering a Allowed Total Error (TEa)<15%. Patients` cc median 83 (4-945) ng/mL. The comparison between ACL TOP 300 and 500 coagulometers showed equivalent results by using the alternative method comparison test. ECT vs. DTI acceptable by Deming`s regression (slope 1.143, Y insert -19.33). cc vs. APTT, TP and DRVVTC nonlinear and moderate correlations with plateau reached at cc >350 ng/mL, r2 0.59, 0.66 y 0.59, respectively. TT is extremely prolonged at cc >50 ng/mL. In

conclusion:

DTI showed a good analytical performance in both coagulometers. ECT showed comparable results to DTI. We verified that dabigatran cc presented moderate correlations with PT, APTT and DRVVTC, and that these tests could only qualitative estimate cc >50 ng/mL.
RESUMO
Os objetivos do trabalho foram verificar a qualidade analítica do ensaio tempo de trombina diluído (DTI) para medição da concentração plasmática (cc) de dabigatrana, comparando dois coagulômetros de detecção foto-óptica, comparar os resultados com o tempo de Ecarina (ECT) e correlacionar as cc com os testes básicos de coagulação Tempo de protrombina (TP), APTT e Tempo de trombina (TT), e tempo de veneno de víbora de Russell com fosfolipídios concentrados (DRVVTC). Foram tomadas 43 amostras de plasma no vale (10-14 h. da última toma) de 40 pacientes que recebiam dabigatrana. DTI e ECT apresentaram (%) repetitividade <5,4% e <7,5%, CV interensaio <6% e <9%, respectivamente, no protocolo EP15A2, aceitáveis para um Erro Total permitido (TEa) <15%. Cc medidas em pacientes mediana 83 (4-945) ng/mL. Comparação de equipamentos ACL TOP 300 e 500 resultados equivalentes por procedimento alternativo de comparação de métodos. Comparação ECT vs. DTI satisfatória por regressão de Deming (pendente 1,143, ordenada à origem -19,33). Correlações cc vs. APTT, TP e DRVVTC moderadas e não lineares tendendo a plateau a cc>350 ng/mL, r2 0,59; 0,66 e 0,59, respectivamente. O TT é extremamente sensível >120 s a cc 50 ng/mL. DTI apresentou um bom desempenho analítico e permitiu quantificar dabigatrana plasmática a cc baixas e altas em ambos os equipamentos utilizados. ECT apresentou resultados comparáveis com DTI. Verifica-se uma correlação moderada entre cc de dabigatrana e os testes clássicos e DRVVTC, podendo ser estimadores de cc a partir dos 50 ng/mL.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Prothrombin Time / Thrombin / Dabigatran Type of study: Practice guideline / Qualitative research Limits: Female / Humans / Male Language: Spanish Journal: Acta bioquím. clín. latinoam Journal subject: Bioqu¡mica / Qu¡mica Cl¡nica Year: 2016 Type: Article Affiliation country: Argentina Institution/Affiliation country: Hospital Británico de Buenos Aires/AR / Hospital Italiano de Buenos Aires/AR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Prothrombin Time / Thrombin / Dabigatran Type of study: Practice guideline / Qualitative research Limits: Female / Humans / Male Language: Spanish Journal: Acta bioquím. clín. latinoam Journal subject: Bioqu¡mica / Qu¡mica Cl¡nica Year: 2016 Type: Article Affiliation country: Argentina Institution/Affiliation country: Hospital Británico de Buenos Aires/AR / Hospital Italiano de Buenos Aires/AR