Falsely increased bispectral index values by convective air warming system during kidney transplantation
Pakistan Journal of Medical Sciences. 2016; 32 (3): 793-795
in En | IMEMR | ID: emr-182988
Responsible library: EMRO
ABSTRACT
Bispectral index [BIS] is a reliable parameter for measuring depth of hypnotic level during anesthesia. Convective air warming system is an effective equipment to maintain normothermia during operation. We report falsely elevated BIS value due to convective air warming system while undergoing kidney transplantation