Medidas do strain longitudinal endocárdico e epicárdico pela técnica do Xstrain®: existem diferenças entre seus valores? / Measures of endocardial and epicardial longitudinal strain by the tchnique of Xstrain®: are there differences between their values? / Medidas del strain longitudinal endocárdico y epicárdico por la técnica del Xstrain®: ¿existen diferencias entre sus valores?
Linhares, Renata Rejane; Paula, Ane Caroline Marques de; Peixoto, Luciana Braz; Monaco, Claudia G; Gil, Manuel A; Ferreira, Luiz Darcy C; Silva, Carlos E. S; Ortiz, Juarez.
Affiliation
  • Linhares, Renata Rejane; OMNI-CCNI Medicina Diagnóstica. São Paulo. BR
  • Paula, Ane Caroline Marques de; DASA - Diagnósticos da América. São Paulo. BR
  • Peixoto, Luciana Braz; DASA - Diagnósticos da América. São Paulo. BR
  • Monaco, Claudia G; OMNI-CCNI Medicina Diagnóstica. São Paulo. BR
  • Gil, Manuel A; OMNI-CCNI Medicina Diagnóstica. São Paulo. BR
  • Ferreira, Luiz Darcy C; OMNI-CCNI Medicina Diagnóstica. São Paulo. BR
  • Silva, Carlos E. S; OMNI-CCNI Medicina Diagnóstica. São Paulo. BR
  • Ortiz, Juarez; OMNI-CCNI Medicina Diagnóstica. São Paulo. BR
ABC., imagem cardiovasc ; 27(1): 2-6, jan.-mar. 2014. ilus
Article in En, Es, Pt | LILACS | ID: lil-705183
Responsible library: BR44.1
RESUMO
Fundamento A ecocardiografia pode avaliar a função ventricular de diversas maneiras pela medida da fração de ejeção ventricular, pela análise visual da contratilidade, ou pela quantificação da deformação miocárdica. Strain é a variável preferida para a descrição da função local.

Objetivo:

Comparar os valores do strain longitudinal endocárdico e epicárdico do ventrículo esquerdo pela técnica do Xstrain® em voluntários saudáveis. Material e

métodos:

Foram avaliados 32 pacientes hígidos com idade média de 31,6 ± 9,8 anos, sendo 18 masculinos (56 por cento). Utilizamos o ecocardiógrafo modelo MyLab60® da Esaote (Firenze, Itália) com sonda multifrequencial (1,5 MHz a 2,6 MHz), e programa de computador para realização das novas técnicas ecocardiográficas (MyLab Desk, v 8.0, Esaote). Mediu-se o strain nas camadas endocárdica e epicárdica aos cortes apicais de 2, 3 e 4 câmaras.

Resultados:

Observou-se diferença estatística entre os valores de strain longitudinal endocárdico e strain longitudinal epicárdico em todos os segmentos estudados (p < 0,01).

Conclusão:

Os valores do strain longitudinal são diferentes nas camadas endocárdica e epicárdica do ventrículo esquerdo.
ABSTRACT

Background:

Echocardiography can assess ventricular function in several ways by measurement the ventricular ejection fraction, by visual analysis of contractility, or by the quantification of myocardial deformity. Strain is the preferred variable for the description of local function.

Objective:

To compare the values of the endocardial and epicardial longitudinal strain of the left ventricle by the technique of Xstrain® in healthy volunteers. Material and

Methods:

Thirty-two healthy patients were assessed with a mean age of 31.6 ± 9.8 years, being 18 men (56 percent). We used echocardiograph model MyLab60® of Esaote (Firenze, Italy) with multifrequency probe (1.5 MHz to 2.6 MHz), and computer program for the performance of new echocardiography techniques (MyLab Desk, v 8.0, Esaote). The strain was measured in the endocardiac and endocardiac layers to the apical 2, 3 and 4 chamber views.

Results:

There was a statistical difference between the values of endocardial and epicardial longitudinal strain and longitudinal epicardiac strain in all segments studied (p < 0.01).

Conclusion:

The values of longitudinal strain are different in endocardial and epicardial layers of the left ventricle.
RESUMEN

Background:

Echocardiography can assess ventricular function in several ways by measurement the ventricular ejection fraction, by visual analysis of contractility, or by the quantification of myocardial deformity. Strain is the preferred variable for the description of local function.

Objective:

To compare the values of the endocardial and epicardial longitudinal strain of the left ventricle by the technique of Xstrain® in healthy volunteers. Material and

Methods:

Thirty-two healthy patients were assessed with a mean age of 31.6 ± 9.8 years, being 18 men (56%). We used echocardiograph model MyLab60® of Esaote (Firenze, Italy) with multifrequency probe (1.5 MHz to 2.6 MHz), and computer program for the performance of new echocardiography techniques (MyLab Desk, v 8.0, Esaote). The strain was measured in the endocardiac and endocardiac layers to the apical 2, 3 and 4 chamber views.

Results:

There was a statistical difference between the values of endocardial and epicardial longitudinal strain and longitudinal epicardiac strain in all segments studied (p < 0.01).

Conclusion:

The values of longitudinal strain are different in endocardial and epicardial layers of the left ventricle. (Arq Bras Cardiolimagem cardiovasc. 2014;27(1)2-6)
Subject(s)
Key words
Full text: 1 Index: LILACS Main subject: Stroke Volume / Echocardiography / Ventricular Dysfunction / Myocardial Contraction Type of study: Prognostic_studies / Risk_factors_studies Limits: Adult / Female / Humans / Male Language: En / Es / Pt Journal: ABC., imagem cardiovasc Journal subject: CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Year: 2014 Type: Article
Full text: 1 Index: LILACS Main subject: Stroke Volume / Echocardiography / Ventricular Dysfunction / Myocardial Contraction Type of study: Prognostic_studies / Risk_factors_studies Limits: Adult / Female / Humans / Male Language: En / Es / Pt Journal: ABC., imagem cardiovasc Journal subject: CARDIOLOGIA / DIAGNOSTICO POR IMAGEM Year: 2014 Type: Article