Your browser doesn't support javascript.
loading
Magnets and implantable cardioverter defibrillators: What's the problem.
Ann Card Anaesth ; 2013 Jan; 16(1): 54-57
Article in English | IMSEAR | ID: sea-145394
ABSTRACT
A growing number of surgical patients present to the operating room with implantable cardioverter defibrillators (ICD). Peri-operative care of these patients dictates that ICD function be suspended for many surgical procedures to avoid inappropriate, and possibly harmful, ICD therapy triggered by electromagnetic interference (EMI). An alternative to reprogramming the ICD is the use of a magnet to temporarily suspend its function. However, this approach is not without complications. We report a case where magnet use failed to inhibit ICD sensing of EMI, and a shock was delivered to the patient. Measures to decrease EMI, controversies regarding magnet use, and expert recommendations are discussed.
Subject(s)

Full text: Available Index: IMSEAR (South-East Asia) Main subject: Ventricular Fibrillation / Aged / Humans / Male / Defibrillators, Implantable / Perioperative Care / Defibrillators / Electromagnetic Fields / Perioperative Period / Cardiac Resynchronization Therapy Devices Type of study: Practice guideline Language: English Journal: Ann Card Anaesth Year: 2013 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: IMSEAR (South-East Asia) Main subject: Ventricular Fibrillation / Aged / Humans / Male / Defibrillators, Implantable / Perioperative Care / Defibrillators / Electromagnetic Fields / Perioperative Period / Cardiac Resynchronization Therapy Devices Type of study: Practice guideline Language: English Journal: Ann Card Anaesth Year: 2013 Type: Article