Your browser doesn't support javascript.
loading
Effect of Clinician-directed Technical Specifications on Entrance Skin Doses in Neonates
Indian Pediatr ; 2023 Jul; 60(7): 537-540
Article | IMSEAR | ID: sea-225435
ABSTRACT

Objectives:

To compare the entrance skin doses (ESD) before and after implementation of a radiation safety policy in neonates (RSN), which focused on clinician-directed technical specifications on the digital X-ray machine.

Methods:

Prospective observations included two sets of X-rays Before (BRSN) and after (ARSN) implementation of RSN (documented indication for Xray/expected posttest findings, settings of 40 kVp, 0.5 mAs, film-focus distance 100 cm, gonadal-shield, optimal collimation, and post-shoot image-enhancement).

Results:

33 and 32 X-rays were analyzed in respective groups. Mean (SD) of calculated and machine-quantified ESD (µGy/m2) was higher in BRSN group as compared to ARSN group (P <0.001). All ARSN X-rays were interpretable for expected post-test findings.

Conclusion:

Clinicians’ cognizance of ability to make consequential bedside technical specifications, can reduce ESD without affecting interpretability. These single observations could have a larger impact in sick neonates, where multiple X-rays are done.

Full text: Available Index: IMSEAR (South-East Asia) Journal: Indian Pediatr Year: 2023 Type: Article