Unidirectional valve malfunction by the breakage or malposition of disc: two cases report / 대한마취과학회지
Article in En | WPRIM | ID: wpr-100098
Responsible library: WPRO
ABSTRACT
Malfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.
Subject(s)
Key words
Full text: 1 Index: WPRIM Main subject: Respiration / Barotrauma / Ventilation / Capnography / Hypercapnia Language: En Journal: Korean Journal of Anesthesiology Year: 2013 Type: Article