Your browser doesn't support javascript.
loading
Comparison of two luminescence detection methods for staphylococcal enterotoxin C content in simulated milk samples / 细胞与分子免疫学杂志
Article in Zh | WPRIM | ID: wpr-1009459
Responsible library: WPRO
ABSTRACT
Objective To compare the sensitivity and accuracy of amplified luminescent proximity homogeneous assay linked immunosorbent assay (AlphaLISA) and magnetic particles-based chemiluminescence immunoassay (MP-CLIA) for detection of staphylococcal enterotoxin C (SEC) in the simulated milk samples. Methods The AlphaLISA was constructed using goat anti-SEC polyclonal antibody-coupled receptor microspheres, biotin-labeled SEC monoclonal antibody and streptavidin-coupled donor microspheres. The MP-CLIA was constructed using goat anti-SEC polyclonal antibody conjugated alkaline phosphatase, biotin-labeled anti-SEC monoclonal antibody and streptavidin conjugated magnetic beads. Results The sensitivity of AlphaLISA to detect SEC content in simulated milk samples was 4.04 ng/L, and the coefficient of variation (CV) was 1.98%~9.82%. The sensitivity of MP-CLIA was 108.19 ng/L and CV was 4.63%~20.40%. Conclusion Compared with MP-CLIA, AlphaLISA is more sensitive and accurate to detecting SEC.
Subject(s)
Full text: 1 Index: WPRIM Main subject: Biotin / Goats / Immunoassay / Streptavidin / Milk / Luminescence / Antibodies, Monoclonal Limits: Animals Language: Zh Journal: Chinese Journal of Cellular and Molecular Immunology Year: 2023 Type: Article
Full text: 1 Index: WPRIM Main subject: Biotin / Goats / Immunoassay / Streptavidin / Milk / Luminescence / Antibodies, Monoclonal Limits: Animals Language: Zh Journal: Chinese Journal of Cellular and Molecular Immunology Year: 2023 Type: Article