Feasibility of acceptance of multiple accelerators using Elekta AGL standard procedures / 中华放射肿瘤学杂志
Article in Zh | WPRIM | ID: wpr-1027497
Responsible library: WPRO
ABSTRACT

Objective:

To verify the feasibility of using Elekta accelerated go live (AGL) standard process for the acceptance of multiple accelerators.

Methods:

The beams of three accelerators were adjusted by PTW Beamscan three-dimensional water tank to reach the AGL standard. Dose verification was performed for three accelerators that met AGL standards. A simple field test example from Cancer Hospital Chinese Academy of Medical Sciences was used to compare the MapCheck 3 surface dose measurement results with the surface dose calculated by the same accelerator model. Images of 10 patients including head and neck, esophagus, breast, lung and rectum were randomly selected. volumetric-modulated arc therapy (VMAT) and intensity modulated radiation therapy (IMRT) treatment techniques were used for planning design, and the measured dose of ArcCheck was compared with the planned dose calculated by the same accelerator model. One-way ANOVA was used to statistically analyze the passing rates of two-dimensional and three-dimensional dose verification.

Results:

The 6 MV X-ray percentage depth dose at 10 cm underwater (PDD 10) of three accelerators was 67.45%, 67.36%, 67.47%, and the maximum deviation between the three accelerators was 0.11%. The 6 MV flattenting filter free (FFF) mode X-ray PDD 10 was 67.33%, 67.20%, 67.20%, and the maximum deviation between the three accelerators was 0.13%. All required discrete point doses on each energy 30 cm×30 cm Profile spindle of the three accelerator X-rays deviated less than ±1% from the standard data. Absolute γ analysis was performed on the results of MapCheck 3 two-dimensional dose matrix validation. Under the 10% threshold of 2 mm/3% standard, the average passing rate of the test cases in Cancer Hospital Chinese Academy of Medical Sciences was above 99%, and the difference was not statistically significant ( P>0.05). Absolute γ analysis was performed on the ArcCheck verification results. Under the 10% threshold, the pass rate of 2 mm/3% was all above 95%, the maximum average passing rate of the three accelerators with different energy and different treatment techniques was 0.28% (6 MV, VMAT), 0.19%(6 MV FFF, VMAT), 0.56% (6 MV, IMRT) and 0.05% (6 MV FFF, IMRT), and the difference was not statistically significant ( P>0.05).

Conclusion:

Compared with traditional accelerator acceptance process, the acceptance time of each accelerator is shortened by 4-6 weeks by using the AGL standard process, and the radiotherapy plan of patients can be interchangeably executed among different accelerators.
Key words
Full text: 1 Index: WPRIM Language: Zh Journal: Chinese Journal of Radiation Oncology Year: 2024 Type: Article
Full text: 1 Index: WPRIM Language: Zh Journal: Chinese Journal of Radiation Oncology Year: 2024 Type: Article