Research on the feasibility of a magnetic-coupling-driven axial flow blood pump / 生物医学工程学杂志
Journal of Biomedical Engineering
; (6): 131-133, 2004.
Article
in Zh
| WPRIM
| ID: wpr-291166
Responsible library:
WPRO
ABSTRACT
A new-designed axial flow blood pump, dived by magnetic coupling and using internal hollow brushless DC motor and inlet and outlet in line with impeller, was tested in mimic circuit. The results showed good performance of the new pump and indicated that its hydrodynamic characteristic can meet the demands of clinical extracorporeal circulation and auxiliary circulation.
Full text:
1
Index:
WPRIM
Main subject:
Assisted Circulation
/
Equipment Design
/
Extracorporeal Circulation
/
Hydrodynamics
/
Magnetics
Language:
Zh
Journal:
Journal of Biomedical Engineering
Year:
2004
Type: