Your browser doesn't support javascript.
loading
Fabrication and Optimization of a Fiber-optic Dosimeter for Proton Beam Therapy Dosimetry / 의학물리
Article in Ko | WPRIM | ID: wpr-30107
Responsible library: WPRO
ABSTRACT
In this study, we have fabricated a fiber-optic dosimeter for a proton beam therapy dosimetry. We have measured scintillating lights with the various kinds of organic scintillators and selected the BCF-12 as a sensor-tip material due to its highest light output and peak/plateau ratio. To determine the optimum diameter of BCF-12, we have measured scintillating lights according to the energy losses of proton beams in a water phantom. Also, we determined the adequate length of organic scintillator by measuring scintillating lights according to the incident angles of proton beam. Using an optimized fiber-optic dosimeter, we have measured scintillating lights according to the dose rates and monitor units of proton accelerator.
Subject(s)
Key words
Full text: 1 Index: WPRIM Main subject: Organothiophosphorus Compounds / Protons / Water / Optical Fibers / Proton Therapy / Light Language: Ko Journal: Korean Journal of Medical Physics Year: 2010 Type: Article
Full text: 1 Index: WPRIM Main subject: Organothiophosphorus Compounds / Protons / Water / Optical Fibers / Proton Therapy / Light Language: Ko Journal: Korean Journal of Medical Physics Year: 2010 Type: Article