Programmable multichannel electrophysiological stimulator based on MCU and CPLD platform / 中国医疗器械杂志
Article in Zh | WPRIM | ID: wpr-323227
Responsible library: WPRO
ABSTRACT
According to the instruction of the operator, the designed programmable multichannel stimulator will deliver stimuli with an assigned amplitude and a proper width to the electrode-array in various modes. Each stimulation time is controlled by the internal timer of the MCU, while the stimulated electrodes' spatial parameter is controlled by the CPLD. Having abandoned conventional isolation using the transformer, we have adapted optical coupling and stable float ground connection to achieve the safety standard required by physiological measurements.
Subject(s)
Full text: 1 Index: WPRIM Main subject: Electric Stimulation Therapy / Electric Stimulation / Electrodes, Implanted / Electrophysiology / Methods Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2008 Type: Article