Your browser doesn't support javascript.
loading
A Preliminary Study on Non-invasive Measurement of Intra-abdominal Pressure / 航天医学与医学工程
Article in Zh | WPRIM | ID: wpr-409741
Responsible library: WPRO
ABSTRACT
Objective To present a new method and design of an instrument for measuring intra-abdominal pressure (IAP) non-invasively. Method A pressure sensor (YH-4) and a displacement sensor (sliding rheostat) were assembled into a probe so that they work in a linear mode. When this assembled instrument probe acts on the abdominal wall of a subject, a pressure as called the external abdominal pressure (EAP), and a corresponding displacement were detected. A relationship was established mathematically between the IAP measured by non-invasive and invasive method, and IAP was calculated by EAP measurement indirectly non-invasively. Result The method was testified by animal experiment in rabbits. And the preliminary results indicated that linear relation between EAP and IAP was obtained. Conclusion Feasibility of the new method is validated by animal experiment. It provides scientific evidence for further clinical experiment.
Key words
Full text: 1 Index: WPRIM Language: Zh Journal: Space Medicine & Medical Engineering Year: 2005 Type: Article
Full text: 1 Index: WPRIM Language: Zh Journal: Space Medicine & Medical Engineering Year: 2005 Type: Article