Establishment and analytical validation of JAK2 V617F mutation assay by fluorescence PCR combined with melting curve analysis / 中华检验医学杂志
Chinese Journal of Laboratory Medicine
; (12): 796-800, 2013.
Article
in Zh
| WPRIM
| ID: wpr-442986
Responsible library:
WPRO
ABSTRACT
Objective According to the requirements of the College of American Pathologists (CAP),to validate the laboratory developed test-JAK2 V617F mutation.Methods This is a clinical laboratory diagnosis study.Fifty-seven peripheral blood and bone marrow samples were collected from the BCR/ABL-negative myeloproliferative neoplasms (MPN) patients in Huashan Hospital from Apr 2010 to Dec 2010.JAK2 V617F mutation was detected by fluorescence PCR combined with melting curve analysis.The accuracy,precision,analytical sensitivity,and anti-interference ability of the assay were validated.The Kappa test was used to evaluate accuracy.Coincidence rate was used to evaluate the repeatability.Paired t test was used to evaluate analytical sensitivity.Results There was a close agreement between the reference method(sequencing) and the test method (fluorescence melting curve analysis) (Kappa =0.89).The coincidence rate was 100%.The results of the assay was not affected by lipoprotein (< 24 mmol/L) or bilirubin (<450 μmol/L).The detection limit of the assay was 5%.Conclusion JAK2 V617F mutation assay by fluorescence PCR combined with melting curve analysis could be applied to the clinical laboratory.
Full text:
1
Index:
WPRIM
Language:
Zh
Journal:
Chinese Journal of Laboratory Medicine
Year:
2013
Type:
Article