Your browser doesn't support javascript.
loading
Evaluation of fitness in implant screw as tightening torque in dental laboratory
Article in Ko | WPRIM | ID: wpr-45375
Responsible library: WPRO
ABSTRACT
PURPOSE: The purpose of this study was to measure the tightening torque for dental implant in dental laboratory and to analyze of the effects of different tightening torque. MATERIALS AND METHODS: The tightening torque for dental implant in dental laboratory were measured by digital torque gauge. The length of abutment and analog were measured as tightening torque of manufacturer's instructions and the measured value. And the data were statically analyzed. RESULTS: The mean tightening torque of implant screw in dental laboratory was 1.563 +/- 0.332 Ncm. The external type implant system of total length were showing no significant differences but the internal type implant system had difference significant (P < 0.05) when compared with tightening torque. CONCLUSION: The implant prosthesis should be made under manufacturer's instructions especially as tightening torque of screw. For the fidelity of implant prosthesis, dental technician should learn how to use the torque gauge.
Subject(s)
Key words
Full text: 1 Index: WPRIM Main subject: Prostheses and Implants / Dental Implants / Dental Prosthesis / Torque / Dental Technicians / Laboratories, Dental Limits: Humans Language: Ko Journal: Journal of Dental Rehabilitation and Applied Science Year: 2015 Type: Article
Full text: 1 Index: WPRIM Main subject: Prostheses and Implants / Dental Implants / Dental Prosthesis / Torque / Dental Technicians / Laboratories, Dental Limits: Humans Language: Ko Journal: Journal of Dental Rehabilitation and Applied Science Year: 2015 Type: Article