Content Determination of Geniposide in TCM Callus Convenient Wipes Drug Solution by HPLC / 中国药房
China Pharmacy
; (12): 4701-4703, 2015.
Article
in Zh
| WPRIM
| ID: wpr-500855
Responsible library:
WPRO
ABSTRACT
OBJECTIVE:To establish a method for the content determination of geniposide in TCM callus convenient wipes drug solution. METHODS:HPLC was performed on the column of Agilent Zorbax SB-C18 with mobile phase of acetonitrile-0.1%phosphoric(10:90,V/V)at the flow rate of 1.0 ml/min,column temperature was 30 ℃,detection wavelength was 236 nm and the volume was 10 μl. RESULTS:The linear range of geniposide was 0.104 1-1.041 μg (r=0.999 8);RSDs of precision,stability and reproducibility tests were all lower than 2%;recovery was 99.04%-100.82%(RSD=0.85%,n=6). CONCLUSIONS:The method is simple,accurate,reproducible,and can be used for the content determination of geniposide in TCM callus convenient wipes drug solution.