Trueness Controls of Clinical Laboratory to Assess Bias / 现代检验医学杂志
Journal of Modern Laboratory Medicine
;
(4): 147-149, 2016.
Article
in Chinese
| WPRIM
| ID: wpr-502870
ABSTRACT
Precision controls monitor assay random error (reproducibility),accuracy controls assess total error,both random error and systematic error,and trueness (bias)of an assay represents systematic error.A trueness control should be metro-logical traceable,ideally with a target value determined by use of a reference material or a reference method,without metro-logical traceable to a reference measurement system,patient test results for the same patient from different laboratories may not be comparable.A trueness control should be commutable and its analytical response to a reference method and a routine field method should be equivalent to that of a patient sample.And trueness control values should generally be set at medical decision levels and be prepared by the providers of reference materials and manufactured in the same fashion as a calibrator.