OBJECTIVE:
The absence of collateral ventilation (CV) is crucial for effective bronchoscopic lung volume reduction (BLVR) with an endobronchial valve. Here, we assessed whether CT can predict the Chartis™ results. MATERIALS ANDMETHODS:
This study included 69 patients (mean age 70.9 ± 6.6 years; 66 [95.7%] males) who had undergone CT to assess BLVR eligibility. The Chartis™ system (Pulmonox Inc.) was used to check CV. Experienced thoracic radiologists independently determined the completeness of fissures on volumetric CT images.RESULTS:
The comparison between the visual and quantitative analyses revealed that 5% defect criterion showed good agreement. The Chartis™ assessment was performed for 129 lobes; 11 (19.6%) of 56 lobes with complete fissures on CT showed positive CV, while this rate was significantly higher (40 of 49 lobes, i.e., 81.6%) for lobes with incomplete fissures. The size of the fissure defect did not affect the rate of CV. Of the patients who underwent BLVR, 22 of 24 patients (91.7%) with complete fissures and three of four patients with incomplete fissures (75%) achieved target lobe volume reduction (TLVR).CONCLUSION:
The quantitative analysis of fissure shows that incomplete fissures increased the probability of CV on Chartis™, while the defect size did not affect the overall rates. TLVR could be achieved even in some patients with relatively large fissure defect, if they showed negative CV on Chartis™.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Pneumonectomy
/
Ventilation
/
Pulmonary Disease, Chronic Obstructive
/
Emphysema
/
Cone-Beam Computed Tomography
/
Lung
Type of study:
Prognostic study
Limits:
Humans
Language:
English
Journal:
Korean Journal of Radiology
Year:
2019
Type:
Article