PURPOSE:
The aim was to study the correlation between cuff size and outcome after implantation of an AMS 800 artificial urinary sphincter.METHODS:
A total of 473 male patients with an AMS 800 sphincter implanted between 2012 and 2014 were analyzed in a retrospective multicenter cohort study performed as part of the Central European Debates on Male Incontinence (DOMINO) Project.RESULTS:
Single cuffs were implanted in 54.5% and double cuffs in 45.5% of the patients. The cuffs used had a median circumference of 4.5 cm. Within a median follow of 18 months, urethral erosion occurred in 12.8% of the cases and was associated significantly more often with small cuff sizes (P<0.001). Multivariate analysis showed that, apart from cuff size (P=0.03), prior irradiation (P<0.001) and the penoscrotal approach (P=0.036) were associated with an increased erosion rate. Continence rate tended to be highest with median cuff sizes (4–5.5 cm).CONCLUSIONS:
Apart from irradiation and the penoscrotal approach, small cuff size is a risk factor for urethral erosion. Results are best with cuff sizes of 4.5–5.5 cm.Key words
Full text:
1
Index:
WPRIM
Main subject:
Multivariate Analysis
/
Retrospective Studies
/
Risk Factors
/
Cohort Studies
/
Urinary Sphincter, Artificial
Type of study:
Etiology_studies
/
Incidence_studies
/
Observational_studies
/
Prognostic_studies
Limits:
Humans
/
Male
Language:
En
Journal:
International Neurourology Journal
Year:
2019
Type:
Article