Study on the synchronization of biventricular beats with the control mode of left ventricular assist device / 生物医学工程学杂志
Article in Zh | WPRIM | ID: wpr-879251
Responsible library: WPRO
ABSTRACT
Right ventricular (RV) failure has become a deadly complication of left ventricular assist device (LVAD) implantation, for which desynchrony in bi-ventricular pulse resulting from a LVAD is among the important factor. This paper investigated how different control modes affect the synchronization of pulse between LV (left ventricular) and RV by numerical method. The numerical results showed that the systolic duration between LV and RV did not significantly differ at baseline (LVAD off and cannula clamped) (48.52%
Subject(s)
Key words
Full text: 1 Index: WPRIM Main subject: Systole / Heart-Assist Devices / Ventricular Function, Right / Ventricular Dysfunction, Right / Heart Failure Limits: Humans Language: Zh Journal: Journal of Biomedical Engineering Year: 2021 Type: Article