Evaluation on Surgical Robotic Pose Repeatability Based on Laser Tracking Measurement / 中国医疗器械杂志
Article in Zh | WPRIM | ID: wpr-928865
Responsible library: WPRO
ABSTRACT
To provide accurate information for registration and safety evaluation of surgical robot, the pose repeatability measurement method was proposed. According to the terminal instrument of the master-slave surgical robot (such as high-frequency electric knife, ultrasonic knife), a suitable target ball fixture was designed. The node data at 10%, 50% and 100% rated speed were measured respectively. Through data analysis, the pose repeatability property of the tested samples at different speeds was obtained. It has high applicability and repeatability, and can meet the requirements of data traceability and registration testing.
Subject(s)
Key words
Full text: 1 Index: WPRIM Main subject: Robotics / Surgery, Computer-Assisted / Equipment Design / Robotic Surgical Procedures / Lasers Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2022 Type: Article
Full text: 1 Index: WPRIM Main subject: Robotics / Surgery, Computer-Assisted / Equipment Design / Robotic Surgical Procedures / Lasers Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2022 Type: Article