ABSTRACT
OBJECTIVE@#To optimize the maintenance quality management of MRI equipment and ensure the quality and safety of its clinical use.@*METHODS@#The data of failure time and repair time of a MRI equipment in three years were collected by magnetic resonance repair report system, and then the reliability, availability and maintainability(RAM) were studied and analyzed.@*RESULTS@#The results of reliability analysis showed that the communication module was the key subsystem of the MRI equipment. The results of usability analysis showed that RF module was a key subsystem of MRI equipment. Maintainability results showed that the proportion of the MRI equipment not fully utilized due to maintenance-related problems was 2.58%. In order to improve the availability of MRI equipment, the maintenance time of MRI equipment should be shortened.@*CONCLUSIONS@#RAM-based analysis of MRI equipment can help hospital equipment managers to carry out the work of operation optimization, maintenance strategy formulation and safety management of MRI equipment.
Subject(s)
Key words
Full text: 1 Index: WPRIM Main subject: Magnetic Resonance Imaging / Reproducibility of Results / Safety Management / Equipment and Supplies, Hospital / Equipment Safety Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2022 Type: Article
Full text: 1 Index: WPRIM Main subject: Magnetic Resonance Imaging / Reproducibility of Results / Safety Management / Equipment and Supplies, Hospital / Equipment Safety Language: Zh Journal: Chinese Journal of Medical Instrumentation Year: 2022 Type: Article