Your browser doesn't support javascript.
loading
Determination of methylmercury in urine by direct mercury analyzer / 中华劳动卫生职业病杂志
Chinese Journal of Industrial Hygiene and Occupational Diseases ; (12): 304-306, 2023.
Article in Chinese | WPRIM | ID: wpr-986033
ABSTRACT

Objective:

To develop asolvent extraction-direct mercury analyzer method for determination of methylmercury in urine.

Methods:

After the urinehydrolyzesd by hydrobromic acid, methylmercury was extracted by tolueneand reverse-extracted from L-cysteine solution, it was then detectedbydirect mercuryanalyzer.

Results:

The linear range was 0.2-50.0 μg/L, and the related coefficient was 0.9999. The relative standard deviations (RSD) within the group were 5.04%-6.64%, and the RSD between the group were 5.65%-8.11 %. The average recovery efficiencies were 85.4%-95.5%. The detection limitation was 0.0482 μg/L and the quantification concentrations was 0.1607 μg/L.

Conclusion:

The method, which has low detection limit, high sensitivity, easy to operate, is stability for the determination of methylmercury in urine.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Mercury / Methylmercury Compounds Language: Chinese Journal: Chinese Journal of Industrial Hygiene and Occupational Diseases Year: 2023 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Mercury / Methylmercury Compounds Language: Chinese Journal: Chinese Journal of Industrial Hygiene and Occupational Diseases Year: 2023 Type: Article